AD-4411
14
7.3. Actual load calibration
Refer to "6 Operation mode" to move to actual load calibration mode,
then press the [F2/ENTER] key.
7.3.1. Zero calibration
Press the [F2/ENTER] key.
If the [F1/
↓
] key is pressed, the indicator skips zero calibration and
enters span calibration.
The current load cell input signal (mV/V) is displayed.
Check that the STABLE status is ON.
Press the [F2/ENTER] key to execute zero calibration.
If zero calibration fails, "C ErrX" is displayed.
Refer to "7.3.3 Error in actual load calibration" for error details.
If zero calibration is successful, "PASS" is displayed, and zero
calibration is completed.
Press the [F2/ENTER] key to enter span calibration.
Or press the [MODE/ESC] key four times to return to measurement
mode.
× 4