7.3.2. Span calibration
Press the [F2/ENTER] key.
If the [F1/
↓
] key is pressed, the indicator skips span calibration and
enters zero calibration.
Set a calibration weight value by the following key operations
[ZERO/
←
] key: Moves the blinking digit to the left.
[TARE/
↑
] key: Increases the blinking digit by one.
[F1/
↓
] key: Decreases the blinking digit by one
[F2/ENTER] key: Confirm the setting value.
The current load cell input signal (mV/V) is displayed.
Place the calibration weight or apply load to the load cell.
Check that STABLE status is ON.
Press the [F2/ENTER] to execute span calibration
If span calibration fails, "C ErrX" is displayed.
Refer to "7.3.3 Error in actual load calibration" for error details.
If span calibration is successful, "PASS" is displayed, and span
calibration is completed.
Press the [MODE/ESC] key four times to return to measurement mode.
Or press the [F2/ENTER] to enter zero calibration again.