7.3.3. Error in actual load calibration
Error code
Description
The load cell signal voltage at zero calibration exceeds 7.0 mV/V.
Check that the load cell is connected correctly.
The load cell signal voltage at zero calibration is less than -7.0 mV/V.
Check that the load cell is connected correctly.
The calibration weight value for span calibration exceeds the maximum capacity.
Set the calibration weight value to the maximum capacity or less.
The calibration weight value for span calibration is below the scale interval.
Set the calibration weight value to the scale interval or more.
The load cell input signal at span calibration is smaller one at zero calibration.
Check that the load cell is connected correctly.
The load cell signal voltage at span calibration exceeds 7.0 mV/V.
Check that the load cell is connected correctly.