8.7. Center-zero detection
The center of zero is detected and ZERO status turns ON when the gross value is within 1/4 of the minimum
division.
8.8. Stability detection
Stability is detected and the stable status is ON when the stability detection time has passed while the
measured value is maintained within the stability detection width.
Set the following to use this function.
・
Set the stability detection time (CALF08) to a value other than 0.0s (default =1.0s).
・
Set the stability detection width (CALF09) to a value other than 0: Disable (default = 2d).
8.9. High resolution display selection
Press a key to which the high resolution display selection function is assigned to switch the displayed value
between high resolution display (10 times higher) and normal display.
Set the following to use this function.
・
Set the function of [F1] / [F2] key (FncF05 / Fnc06) to 4: high resolution display selection (default = 0: None).
8.10. Comparator
The comparison value and the measured value are compared, and then Hi / OK / Lo is detected when the
following conditions are met.
Hi
Measured value
>
Upper limit value
OK
Upper limit value
≧
Measured value
≧
Lower limit value
Lo
Lower limit value
>
Measured value
Set the following to use this function.
・
Upper limit value (FncF11). (Default = 10)
・
Lower limit value (FncF12). (Default = 10)
・
Select the target for comparison (FncF13) from 1: Gross (default) / 2: Net.