45
AD-4411
15.1. Calibration function list
Modbus
Command
Setting item
Setting value
Default
User
setting
Read Write
400101 1101
10101 Unit
0: None / 1: g / 2: kg / 3: t
2
400103 1103
10103 Decimal point position
0:
0
(No decimal point)
1:
0.0
2:
0.00
3: 0.000
4: 0.0000
5: 0.00000
0
400105 1105
10105 Minimum division
1: 1 d
2: 2 d
3: 5 d
4: 10 d
5: 20 d
6: 50 d
1
400107 1107
10107 Maximum capacity
1 to 999999
999999
400109 1109
10109 Zero setting range
0 to 100 %
100
400111 1111
10111 Zero tracking time
0.0 to 5.0 s *
0.0
400113 1113
10113 Zero tracking width
0: Disable
1: 0.5 d
2: 1.0 d
3: 1.5 d
4: 2.0 d
5: 2.5 d
6: 3.0 d
7: 3.5 d
8: 4.0 d
9: 4.5 d
0
400115 1115
10115 Stability detection time
0.0 to 9.9 s *
1.0
400117 1117
10117 Stability detection width
0 to 100 d
2
400119 1119
10119 Zero-setting at unstable
0: Disable / 1: Enable
1
400121 1121
10121 Taring at unstable
0: Disable / 1: Enable
1
400123 1123
10123 Taring at the negative gross
0: Disable / 1: Enable
1
400125 1125
10125 Zero clear
0: Disable / 1: Enable
1
400127 1127
10127 Power on zero
0: Disable / 1: Enable
0
400129 1129 10129 Condition of negative overload
0: Gross < -(Maximum ca 8d)
1: Gross < -19d
0
400131 1131
10131 NTEP
0: Disable / 1: Enable
0
400133 1133
10133 Load cell connection type
0: 4-wire / 1: 6-wire
1
* When accessed via Ethernet/Modbus, the decimal point position is ignored.
15.2. Digital calibration function list
Modbus
Command
Setting item
Setting value
Default
User
setting
Read Write
401501 2501
11501 Load cell input signal at zero calibration
-7.00000 to 7.00000 mV/V *
0.00000
401503 2503
11503 Load cell input signal
(at span calibration – at zero calibration)
0.00001 to 7.00000 mV/V *
2.00000
401505 2505
11505 Load value at span calibration
1 to 999999
20000
* When accessed via Ethernet/ Modbus, the decimal point position is ignored.