10
4
-
2
Function Table
Table 6
Class Item
Parameter
Description
0
Set 0
When connected to EK-G, EW-G,
FC-
i
, FC-S
i
, FG, FP, FS, FS-KL
HV-G, HV-WP, HW-G or HW-WP.
Disables the AD-8922 keys for use
as a remote display.
1
Set 1
When connected to ET-W, GH, GP,
GX, GX-K.
2
Set 2
When connected to a GR series
balance.
3
Set 3
When connected to GF, GF-K,
AD-4212, HR, HR-
i
. EK-H.
C5et
*
Command set
4
Set 4
When connected to an EK-
i
/EW-
i
series balance.
0
Through mode
Always outputs the data received by
the D-Sub 9-pin connector, to the
DIN 8-pin connector.
1
Key mode 1
Outputs the latest data received by
the D-Sub 9-pin connector, to the
DIN 8-pin connector, when the
AD-8922 PRINT key is pressed.
out
Output mode
2
Key mode 2
Outputs the latest stable data
received by the D-Sub 9-pin
connector, to the DIN 8-pin
connector, when the AD-8922
PRINT key is pressed.
0
Bar display
Bar display if nothing has been
received for two seconds or more.
Hold
Timeout
1
Hold display
Displays the previous data received
if nothing has been received for two
seconds or more, and turns on the
hold indicator.
0
Not used
No buzzer upon data receipt.
1
At intervals
Sounds buzzer when the data is
received with an interval of two
seconds or more.
fnc
Environment
Display
bep
Data receipt
buzzer
2
Used
Sounds buzzer upon each data
receipt.
0
600
bps
1
1200 bps
2
2400 bps
3
4800 bps
4
9600 bps
bp5
Baud rate
5
19200 bps
0
7 bits - EVEN
1
7 bits - ODD
btpr
Data bits
Parity bit
2
8 bits - NONE
0
1 bit
5top
Stop bits
1
2 bits
0
CR/LF
5if
Serial
interface
Crlf
Terminator
1
CR
Choose a parameter appropriate to
the weighing instrument.
When the AD-8121B compact
printer is to be connected, leave the
factory settings of the AD-8922 as is
and set the weighing instrument.
Factory setting
*
See Table 2 in "1-2 Applicable Instruments" to set the command set "
C5et
" of the function setting,
appropriate to the weighing instrument.