26
11
-
3
Setting the Upper and Lower Limit Values
z
Up to 10 set of upper and lower limit values can be stored.
z
Using the stored upper and lower limit values, comparison can be performed easily. To recall the
stored values, press and hold the ON:OFF key and press the MODE key.
11
-
3
-
1
Storing the upper and lower limit values
To store new upper and lower limit values, recall the stored data ("C01" to "C10") and change them.
1. While pressing and holding the ON:OFF key, press the RE-ZERO key to enter the confirmation
mode. The upper limit value data (Comparator number and the upper limit mass (blinking)) of the
comparator number that was selected last.
Displaying
example
2. Select the comparator number using the following keys.
RE-ZERO key
To increase the comparator number by 1.
MODE key
To decrease the comparator number by 1.
Each time the key is pressed, the upper limit value and the lower limit value of the comparator
number selected is displayed alternately. (
C01
HI
⇔
C01
LO
⇔
C02
HI
⇔
C02
LO
⇔…
)
3. Press the SAMPLE key to go to the storing mode to change the stored values.
Digital input mode
SAMPLE key
To select the digit to
change the value.
RE-ZERO key
To change the value of
the digit selected.
MODE key
To move the decimal
point position to the
right by 1 digit.
ON:OFF key+MODE key To switch the polarity.
PRINT
key
To store the new
setting and return to
step 2.
CAL
key
To cancel the new
setting and return to
step 2.
4. Press the CAL key to return to the weighing data
display.