Recalling the upper and lower limit values
The procedure below describes an easy way to recall the upper and lower limit values to be used for
weighing.
1 While pressing and holding the ON:OFF key, press the MODE key to enter the selection mode.
2 The upper limit value last selected with its comparator number appears.
3. Select the comparator number using the following keys.
RE-ZERO key
To increase the comparator number by 1.
MODE key
To decrease the comparator number by 1.
Each time the key is pressed, the upper limit value and the lower limit value of the comparator
number selected is displayed alternately. (
C01
HI
⇔
C01
LO
⇔
C02
HI
⇔
C02
LO
⇔…
)
Only the stored comparator numbers are displayed.
4 Press the PRINT key to confirm the selection and return to the weighing data display with the
selected upper and lower limit values ready for use (In the example shown below, the values of
"
C08
". )
Note:
When no operation is performed in step 4 (after a few seconds of
inactivity), the AD-8922 selects the value currently displayed and returns to the weighing
data display automatically.
To cancel the operation, press the CAL key.