13
Cable connection
CAUTION
Before inserting the power line, make sure that the power to the AD-8923-CC is turned off.
(1) Inserting the power line
Press down the release button on the DC
input terminal using a screwdriver and
insert the power line.
The recommended stripping length for the
power line is 10 mm.
Applicable wire range
Single wire:
φ
1.0 mm (AWG 26) to
φ
1.2 mm (AWG 16)
Twisted wire: 0.3 mm
2
(AWG 22) to 0.75 mm
2
(AWG 20)
Individual wire diameter
φ
0.18 mm or greater
(2) Securing or removing the power line
To secure the power line, return the release button to the initial position using the screwdriver.
The power line will be locked. To remove the power line, press the release button again using
the screwdriver, unlocking the power line.
When the AC adapter is used
Insert the AC adapter plug into the AC adapter
input jack located on the rear of the AD-8923-CC
and insert the AC adapter into an electrical outlet.
3.4. Operation
Displays the data transmitted by the weighing instrument connected.
The keys on the AD-8923-CC can control the weighing instrument. The key operation depends on the
weighing instrument connected. For details, refer to “Table 2” of “
DC input terminal
AD-8923-CC rear
Release button
Screwdriver
Power line
AC adapter input jack
DC input terminal
AC adapter plug
AD-8923-CC rear
Summary of Contents for AD-8923-CC