15
Sensitivity adjustment procedure
Performs sensitivity adjustment with the AD-4212C using an external weight.
Operation
1. Warm up the AD-4212C for 30 minutes or more with nothing on the pan.
2. Press the CAL key.
is displayed.
If you want to cancel sensitivity adjustment, press the CAL key.
The display will return to the weighing mode.
If you want to change the weight value, press the SAMPLE key.
Press the RE-ZERO key to select the weight value,
and press the PRINT key to store it.
is displayed.
3. Confirm that there is nothing on the pan and press the PRINT key.
The AD-4212C measures the zero point. Do not allow vibration or
drafts to affect the AD-4212C. The weight value is displayed.
4. Place a weight, of the weight value displayed, on the pan and
press the PRINT key. The AD-4212C measures the weight.
Do not allow vibration or drafts to affect the AD-4212C.
5.
is displayed. Remove the weight from the pan.
6. The display will automatically return to the weighing mode.
7. Place the weight on the pan and confirm that adjustment was performed
correctly. If not, check the ambient conditions such as drafts or vibration,
and repeat steps 2 through 7.
*
It is also possible to carry out sensitivity adjustment described above using the register of the
CC-Link. For details, refer to “
6.4. Sensitivity adjustment with the AD-4212C/F using the register
”.
Press
Weight
Press
Press
STABLE FAST MID.
RESPONSE
SLOW
CAL
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
Summary of Contents for AD-8923-CC
Page 1: ...1WMPD4002124D...
Page 32: ...32 MEMO...
Page 33: ...33 MEMO...
Page 34: ...34 MEMO...