Numeric values of the remote register
All the values are hexadecimal. Negative values are expressed by the two’s complement.
Decimal
Hexadecimal (32 bits)
-10
FFFFFFF6
-1
FFFFFFFF
0
00000000
1
00000001
10
0000000A
Weight value examples
1.000 will be 1000, thus expressed as 0x000003E8. (RWr0001: 0x0000, RWr0000: 0x03E8)
-1.000 will be -1000, thus expressed as 0xFFFFFC18. (RWr0001: 0xFFFF, RWr0000: 0xFC18)
Prohibited writing in the internally reserved areas
Writing is prohibited in the internally reserved areas.
Writing in the remote output (RY) and the remote register (RWw) of the internally reserved areas
may cause the AD-8923-CC to malfunction.
Values of the remote input (RX) and the remote register (RWr) of the internally reserved areas are
not fixed.
Stable
/
Unstable
RX0007
0
Unstable
1
Stable
Decimal point
RX0008 to RX000A, 3-bit binary notation
RX000A
RX0009
RX0008
Decimal point position
0
0
0
No decimal point
0
0
1
First digit
0
1
0
Second digit
0
1
1
Third digit
1
0
0
Fourth digit
1
0
1
Fifth digit
Decimal point position example
When displaying 1.000, express 3 as a decimal point at the third digit, thus 0x011. (RX000A: 0, RX0009:
1, RX0008: 1)
State flag of sensitivity adjustment *
RX0002 RX0001 RX0000
State
0
0
1
Waiting for zero point input
0
1
0
Waiting for sensitivity adjustment mass value input
0
1
1
Sensitivity adjustment completed
1
0
0
Sensitivity adjustment error