6.4. Sensitivity adjustment with the AD-4212C/F using the register of the CC-Link
The following describes the sensitivity adjustment procedure using the register of the CC-Link when
connecting the AD-8923-CC to the AD-4212C
/
F. (Using an external weight)
* Only AD-8923-CC software versions P2.05 or later are compatible with this function.
* When performing sensitivity adjustment using the key operation, refer to “
” for the details.
Caution on sensitivity adjustment
While the sensitivity adjustment is being carried out, take care to use these instruments in an environment
where they are not affected by vibration, drafts or temperature change.
Sensitivity adjustment
using key operation
State and operation of
the balance
Weighing state
(CAL key input)
Waiting state for
zero input
Confirm that there is nothing
on the weighing pan.
(PRINT key input)
Acquiring zero state
Waiting for
sensitivity adjustment mass
(PRINT key input)
Acquiring sensitivity
adjustment mass
Sensitivity adjustment
complete
Remove the mass
from the weighing pan.
Weighing
Register to be used
When the sensitivity adjustment cannot be carried out, the state flag of sensitivity adjustment becomes as follows.
Enter into waiting
state for the zero
input by RY003:0
1.
Begin to acquire
the mass by RY004:0
1.
Begin to acquire
the zero by RY004:0
1.
AD-8923-CC Master station Master station AD-8923-CC
Sensitivity adjustment error
CAL
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW
STABLE FAST MID.
RESPONSE
SLOW