3.5. Performing Sensitivity Adjustment with the AD-4212C
/
F
The following is the sensitivity adjustment procedure when the AD-4212C
/
F is connected.
(An external weight is used.)
Caution
Do not allow vibration, drafts or temperature change to affect the AD-4212C during sensitivity
adjustment.
Caution on using an external weight
The accuracy of the weight can influence the accuracy of weighing.
Select an appropriate weight as listed below.
A weight of 200 g is provided with the AD-4212C as a standard accessory.
Weighing instrument
Usable weight
AD-4212C-300
AD-4212C-301
50 g, 100 g,
200 g
,
300 g
AD-4212C-600
50 g, 100 g,
200 g
,
300 g, 400 g, 500 g, 600 g
AD-4212C-3000
AD-4212C-3100
50 g, 100 g,
200 g
, 300 g, 500 g, 1000 g, 2000 g, 3000 g
AD-4212C-6000
200 g
, 500 g, 1000 g, 2000 g, 3000 g, 4000 g, 5000 g, 6000 g
AD-4212F-6203D
50 g, 100g, 200 g, 300 g, 500 g, 1000 g,
2000 g
, 3000 g, 4000 g,5000 g, 6000 g
AD-4212F-10202
500 g, 1000 g, 2000 g, 3000 g, 4000 g,
5000 g
, 6000 g ,7000 g, 8000 g, 9000 g, 10000 g
AD-4212F-22001
1000 g, 2000 g, 5000 g,
10000 g
, 20000 g
The weight in bold type: Factory setting
Display
This indicator means “the AD-4212C is measuring sensitivity adjustment
data”. Do not allow vibration, drafts or other external disturbances to
affect AD-4212C while this indicator is displayed.