6.5. Four corner adjustment
The example shown in the figure is when the weighing scale is built using four digital
load cells.
Press [PRINT].
Put the adjustment weight on the 1st digital load cell.
When the MD mark illuminates, press [PRINT] to execute corner
adjustment for 1st digital load cell.
Press [PRINT].
Put the adjustment weight on the 2nd digital load cell.
When the MD mark illuminates, press [PRINT] to execute corner
adjustment for 2st digital load cell.
Execute corner adjustment for all of the load cells connected
by using the same procedure.
After that, the display proceeds as shown in the example at
the right.
lC1adj
lC2adj
4C end
Previous page
1000.0
lC3adj
1000.0
Execute
Execute
For 2 seconds
Next page
Cal 0