background image

 

AD4329A-DLC 

Page 22 

 

6.9. Scale interval and capacity settings for each range 

Set the scale interval and capacity settings for each range by the following key operations. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

         

0.0 

 

Cal in 

     

d-01 

After 2 seconds 

     

d-02 

Update  
the value 

C "10000 

Change 
the value 

C

   

"2000

 

Cap2-05 

Cap2-10 

C

   

"2000

 

C  "5000 

Cap3-20 

Cap3-50 

C "10000 

C "20000 

Scale interval setting 

for 1st range *1 

   

Cal 0 

Zero calibration 

*1) Press [TARE] within 2 seconds after "d-XX appears in the

display, and the scale interval setting for the 2nd or 3rd range

will appear in the display. If not, the capacity setting will

automatically appear in the display. 

Scale interval setting

for 2nd range *1 

Scale interval setting

for 3rd range *1 

Capacity setting   

for 1st range *1 

Capacity setting   

for 2nd range *1 

Capacity setting   

for 3rd range *1 

Move  
the digit 
place 

Change 
the value 

Summary of Contents for AD4329A-DLC

Page 1: ...AD4329A DLC Weighing Indicator INSTRUCTION MANUAL 1WMPD4003952...

Page 2: ...bration 15 6 1 Scale interval setting 17 6 2 Maximum capacity setting 17 6 3 Zero compensation at four corner adjustment 18 6 4 Adjustment weight at four corner adjustment 18 6 5 Four corner adjustmen...

Page 3: ...ng the multi interval settings 50 12 9 3 Confirming the linearity calibration 50 12 9 4 Confirming the gravitational acceleration compensation 1 50 12 9 5 Confirming the gravitational acceleration com...

Page 4: ...ng system with higher lightning resistance performance can be built and electronic corner adjustment can be performed easily and precisely Up to eight digital load cells can be connected in parallel A...

Page 5: ...with oil chemicals or water Please note that securing the indicator to the control panel will provide the indicator outside of the control panel with IP65 protection When installing or removing the i...

Page 6: ...of the minimum division 4 Key switch with sealingcover CAL key Switches to calibration mode OPR STB key Turns the display on off Preset tare key Displays the preset tare value M key Adds the current...

Page 7: ...Fuse holder 4 External control input connector 5 Standard serial output connector 6 RS 232C I F connector 7 Option 02 comparator output slot 2 3 Accessories Name Qty Instruction Manual 1 Connector for...

Page 8: ...ion 1 AC power input Live Unearthed conductive part 2 AC power input Neutral Earthed conductive part 3 Ground terminal The test below is performed for lightning resistance performance of the AD4329A D...

Page 9: ...Load cell input terminal assignment Use an M3 solderless terminal with a width of 6 2 mm or less Terminal No Symbol Description 4 SHILD Shield 5 POW Load cell excitation voltage 6 POW Load cell excit...

Page 10: ...connect the digital load cells in parallel use a junction box for digital load cells Connecting the weighing indicator to four digital load cells is as shown below Recommended cable is AX KO3217 A D...

Page 11: ...furthest from the weighing indicator as is and cut the jumpers JP1 in the other junction boxes using wire cutters Fig 4 shown below is an example connection diagram when using an AD4388 4 and an AD43...

Page 12: ...F function mode f 01 Press CAL to go to CF function mode Select Cf 100 by the following key operations ZERO Move the flashing digit TARE Change the value of the flashing digit PRINT Confirm the displ...

Page 13: ...n setting the serial number set it in the order for corner adjustment starting from Cf 101 Check the serial number marked on each digital load cell For example with the LCCD20 series manufactured by A...

Page 14: ...the following key operations ZERO Move the flashing digit TARE Change the value of the flashing digit PRINT Update to the new value and return to the item selection MODE Cancel the value setting and...

Page 15: ...the following items required for calibration Scale interval resolution setting Maximum capacity setting Four corner adjustment Zero calibration Span calibration Linearity calibration if necessary Gra...

Page 16: ...e interval setting d 01 C 10000 Cal 0 f 10000 Maximum capacity setting Zero calibration Span calibration 4C Cal0 g1 9798 Cap2 02 lnr 1 Linearity calibration Multi interval function Gravitational accel...

Page 17: ...to the next item MODE Skip this item and go to the next item 6 2 Maximum capacity setting Set a maximum capacity by the following key operations TARE Change the value of the flashing digit ZERO Move...

Page 18: ...adjustment and go to the next item 6 4 Adjustment weight at four corner adjustment Press PRINT Set an adjustment weight value for corner adjustment by the following key operations TARE Change the val...

Page 19: ...rner adjustment for 1st digital load cell Press PRINT Put the adjustment weight on the 2nd digital load cell When the MD mark illuminates press PRINT to execute corner adjustment for 2st digital load...

Page 20: ...he flashing digit ZERO Move the flashing digit MODE Skip span calibration and return to scale interval setting Place the calibration weight on the scale When the MD mark illuminates press PRINT to exe...

Page 21: ...g 1st range Interval 0 01kg is selected When the weighing value is between 20 and 50kg 2nd range Interval 0 02kg is selected Display When the weighing value is between 50 and 100kg 3rd range Interval...

Page 22: ...000 C 20000 Scale interval setting for 1st range 1 Cal 0 Zero calibration 1 Press TARE within 2 seconds after d XX appears in the display and the scale interval setting for the 2nd or 3rd range will a...

Page 23: ...point Set the following items to adjust linearity Calibration for compensation point 1 Calibration for compensation point 2 If necessary Calibration for compensation point 3 If necessary Before compe...

Page 24: ...ight by the following key operations TARE Change the value of the flashing digit ZERO Move the flashing digit MODE Delete compensation point 2 and 3 to go to span calibration Place the adjustment weig...

Page 25: ...ing Set the gravitational acceleration for the calibration location by the following key operations The decimal point is not displayed TARE Change the value of the flashing digit ZERO Move the flashin...

Page 26: ...hing range is larger than the next weighing range err 13 Shows that the adjustment weight settings for linearity adjustment don t meet the following conditions The weight setting for lnr 1 The weight...

Page 27: ...lay in weighing mode are as follows Weighing value Positive overload Negative overload Error display Description errAd n The n th digital load cell has a problem in the internal circuit errlC n The n...

Page 28: ...set to zero automatically when the gross value meets the zero tracking condition F 01 7 6 Power on zero After turning on the power the gross value is set to zero automatically when the gross value is...

Page 29: ...ue is set to 0 by the following operations Turn off power Pressing TARE when the gross value is 0d Power on zero Push zero 7 10 Preset tare This subtracts a preset tare from the gross value If you use...

Page 30: ...d 3 20d 4 50d Accumulation execution Manual accumulation When M is pressed with the MD mark illuminated accumulation is executed and the display flashes once Auto accumulation When the MD mark is illu...

Page 31: ...Press MODE 2 After total is displayed the total value is displayed with M mark flashing 3 Press and hold ZERO for 3 seconds or longer 4 After the display flashes once the total value is cleared to 0...

Page 32: ...nufactured by Hosiden is included as an accessory If you use external control inputs select the functions for each pin from the following settings 0 None default 1 ZERO key 2 TARE key 3 PRINT key 4 OP...

Page 33: ...6 715267 connector manufactured by Hosiden is included as an accessory Pin No Description 1 NC 2 NC 3 Current loop output no polarity 4 NC 5 Current loop output no polarity 6 NC 7 NC Shell Frame groun...

Page 34: ...If you connect a printer without receive buffer set the data interval F 33 to 1 2 seconds Select communication mode F 31 from the following settings 0 Stream default Weighing data is output at the di...

Page 35: ...34 5 Numerical value with polarity 0000000 0 SP SP SP SP SP SP Positive overload Weighing data 1 SP SP SP SP SP SP Negative overload kg Kiro gram Unit CF 001 1 SP t Ton 1 It is possible to select the...

Page 36: ...times If you use the comparator output set the following settings Set comparator function F 20 to 1 Enable Default is 0 Disable Select comparator function inhibition condition F 21 from the following...

Page 37: ...ing digit NET GROSS Switch the polarity PRINT Update the value MODE Skip the upper limit value lo will appear on the display for 2 seconds Change the lower limit value by the following key operations...

Page 38: ...panel with the main board 3 Remove two M3 screws on the COMP OUT slot and the blank panel 4 Install the comparator output option to the rear panel with two M3 screws 5 Connect the 5 wire cable between...

Page 39: ...h screws The connector is not provided and must be prepared by the user Pin No Description 1 NC 2 TXD output 3 RXD input 4 NC 5 SG 6 9 NC Communication specifications Signal level RS 232 Baud rate F 4...

Page 40: ...3 to 1 2 seconds Select communication mode F 41 from the following settings 0 Stream default Weighing data is output at the display update rate 1 Auto print When the MD mark is illuminated a weighing...

Page 41: ...0000 0 0 SP SP SP SP SP SP Positive overload Weighing data 1 SP SP SP SP SP SP Negative overload kg Kilograms Unit CF 001 1 SP t Ton 1 it is possible to select the number of digits and decimal point f...

Page 42: ...ross value MN CR LF Switch the display to net value When transmitting the following command and writing data is sent the data will be updated and the command and writing data will be echoed back PT 21...

Page 43: ...CHc1 o U t CHc r 5 CHc f CHc1 C f CHc1 C a l Off mode Weighing mode init Confirmation of the digital load cell Confirmation of the keys Confirmation of the standard serial output Initialization Confir...

Page 44: ...splaying lC4 5n for 2 seconds the serial number is displayed After displaying lC4 Cap for 2 seconds the maximum capacity is displayed After displaying lC4load for 2 seconds the actual load applied to...

Page 45: ...4329A DLC Page 45 12 2 Confirming the keys Press the each key to switch the corresponding o to 1 oooooo 1ooooo o1oooo oo1ooo ooo1oo oooo1o ooooo1 ooo 1oo o1o oo1 2 seconds later CHc1yey 2 seconds late...

Page 46: ...COM the corresponding o to switched to 1 12 4 Confirming the standard serial output Press PRINT to transmit ST G 00000 0kg CR LF Communication specifications depend on settings for F 30 to F 34 Cl Cl...

Page 47: ...munication specifications depend on settings for F 40 to F 47 When receiving data the first four characters are displayed The example at the right is when A123 CR LF is received Communication specific...

Page 48: ...splay the setting value for the currently displayed item 12 8 Confirming the CF function setting values Operate the following keys to select the items to be confirmed TARE Change the value of the flas...

Page 49: ...val Multi interval settings Linearity calibration Gravitational acceleration compensation 1 Confirmation of each setting value Calf CalC Maximum capacity Gravitational acceleration compensation 2 Weig...

Page 50: ...erval settings for 2 seconds 12 9 3 Confirming the linearity calibration Press PRINT to proceed to the next confirmation item after displaying the number of compensation points for linearity calibrati...

Page 51: ...em after displaying the weight value at span calibration for 2 seconds 12 9 7 Confirming the load cell output at zero calibration Press PRINT to proceed to the next confirmation item after displaying...

Page 52: ...to be confirmed TARE Change the load cell to be confirmed PRINT Return to previous display after displaying the compensation value for 2 seconds 12 9 11 Compensation value for four corner adjustments...

Page 53: ...dition to the parameter Initialization of all data Initializes the calibration settings in addition to the parameter and function and sets the weighing indicator to the factory settings Follow the pro...

Page 54: ...t TARE Change the value of the flashing digit PRINT Confirm the displayed item Change the value by the following key operations ZERO Move the flashing digit TARE Change the value of the flashing digit...

Page 55: ...umulation mode and value polarity condition 0 1 2 3 Manual Only plus value Manual Both plus minus value Auto Only plus value Auto Both plus minus value 0 05 Accumulation inhibition range 0 1 2 3 4 0 d...

Page 56: ...11 EXT2 Pin3 0 12 EXT3 Pin4 0 13 EXT4 Pin5 0 14 EXT5 Pin6 0 15 EXT6 Pin7 0 16 External control input function EXT7 Pin8 0 1 2 3 4 5 6 7 8 9 10 11 None ZERO key TARE key PRINT key OPR STB key M key NE...

Page 57: ...d or less or unstable Displayed value is within 5d Displayed value is within 5d or is unstable 0 13 2 4 Standard serial output F Item Setting value Default User setting 30 Output data 0 1 2 3 4 Displa...

Page 58: ...e value Gross Net Tare value 0 41 Communication mode 0 1 2 3 4 Stream Auto print Manual print Command Reserved 0 42 Auto print inhibition range 0 1 5 d or less Within 5 d 0 43 Data interval at F 40 4...

Page 59: ...E Change the value of the flashing digit PRINT Confirm the displayed item Change the value by the following key operations ZERO Move the flashing digit TARE Change the value of the flashing digit PRIN...

Page 60: ...ero 0 1 Disable Enable 1 006 Execution Preset tare 0 1 Enable Disable 1 007 When outputting the net value the preset tare value is output as well 0 1 Disable Enable 1 008 Output on over load and unsta...

Page 61: ...61 CF Item Setting value Default User setting 100 Number of digital load cells 1 to 8 4 101 1st 102 2nd 103 3rd 104 4th 105 5th 106 6th 107 7th 108 Serial number of each digital load cell 8th 0000000...

Page 62: ...ge 62 14 Check the Software Version Press OPR STB While pressing and holding MODE press OPR STB Press MODE The software version will appear Press OPR STB twice to back to weighing mode 0 00 f 0 1 OPR...

Page 63: ...to eight A D digital load cells can be connected in parallel Sampling rate 10 times s Display Weight display 7 digit LED with character height of 13 mm Units kg t selectable Status lamp M MD GROSS NE...

Page 64: ...AD4329A DLC Page 64 External Dimensions Unit mm Panel cut dimensions Panel thickness 1 6 to 3 2...

Page 65: ...44 1235 550485 A D AUSTRALASIA PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited 6 33 817 07331 817 Manhattan Bldg 33 Gukjege...

Reviews: