Linearity adjustment execution
Press [PRESET TARE] in the
calibration weight setting for span calibration.
"Lnr 1" will appear in the display for 2 seconds.
Set the adjustment weight by the following key operations:
[TARE]:
Change the value of the flashing digit.
[ZERO]:
Move the flashing digit.
[MODE]:
Delete compensation point 1, 2, and 3 to go
to span
calibration.
Place the adjustment weight on the scale. When the MD mark illuminates,
press [PRINT] to execute adjustment for the point 1.
"Lnr 2" will appear in the display for 2 seconds.
Set the adjustment weight by the following key operations:
[TARE]:
Change the value of the flashing digit.
[ZERO]:
Move the flashing digit.
[MODE]:
Delete compensation point 2 and 3
to
go to span
calibration.
Place the adjustment weight on the scale. When the MD mark illuminates,
ON, press [PRINT] to execute adjustment for the point 2.
"Lnr 3" will appear for 2 seconds.
Set the adjustment weight by the following key operations:
[TARE]:
Change the value of the flashing digit.
[ZERO]:
Move the flashing digit.
[MODE]:
Delete compensation point 3 to
go to span calibration.
Place the adjustment weight on the scale. When the MD mark illuminates
,
press [PRINT] to execute adjustment for point 3.
l "10000
l "5000
Change Move
f "10000
Change Move
l "2500
Execute
l "10000
Change Move
Execute
l "10000
l "7500
Execute
f "10000