Upper limit and lower limit setting
Press [ON/OFF] in weighing mode.
While pressing and holding [PRINT], press [OPR/STB].
"
Hi
" will appear on the display for 2 seconds.
Change the upper limit value by the following key operations.
The decimal point does not appear in the display.
Example: Input 3000 to set 300.0 for the upper limit
[TARE]:
Move the flashing digit.
[ZERO]:
Change the value of the flashing digit.
[NET/GROSS]:
Switch the polarity.
[PRINT]:
Update the value.
[MODE]:
Skip the upper limit value.
"
lo
" will appear on the display for 2 seconds.
Change the lower limit value by the following key operations.
The decimal point does not appear in the display.
Example: Input 500 to set 50.0 for the lower limit
[TARE]:
Move the flashing digit.
[ZERO]:
Change the value of the flashing digit.
[NET/GROSS]:
Switch the polarity.
[PRINT]:
Update the value
[MODE]:
Skip the lower limit value.
Hi
500"0
"3000
0.0
.
Standby mode
Change Move
Polarity
Update
Skip
lo
100"0
"500
Change Move
Polarity Skip
Update
Hi
After 2 seconds
After 2 seconds