2-8
2.4 Indication upon Balancing
Pressing BAL knob
④
executes wire disconnection check and wire length adjustment, thus
automatically realizing both resistance balancing and capacitance balancing. While the wire
disconnection check and wire length adjustment functions are effective (ON), the following
indications appear. If there is wire disconnection or short, the problematic location is indicated
repeatedly on digital monitor
③
. The indication continues until the BAL knob is pressed for five
seconds or longer or the power of amplifier is turned off. Make repair following the indication. When
wire disconnection check is not effective (OFF), the portions enclosed by broken line are omitted,
executing auto-balancing. For how to set these functions to ON or OFF, refer to page 3-10 How to
Switch Special Function Setting.
Cable Length Compensation Check :
ON
Figure.2-5 Block
diagram for balance
operation
or
Push BAL. knob.
Display bridge check
Checking AB line
Checking BC line
Checking CD line
Checking DA line
No disconnection or short
Disconnection or short
Repeatedly showing
until BAL knob being
pressed
Example of Indication: When there are
disconnections at AB and DA on the bridge
Disconnection in
AB on bridge
Disconnection in DA on bridge
Disconnection on wire A
No disconnection or short : Good
Bridge check : ON
Indicating dropout rate
Cable length
Compensation check
Auto Balance
Summary of Contents for AS2503
Page 1: ...AS2503 AS2603 DC BRIDGE SYSTEM DYNAMIC STRAIN AMPLIFIERS INSTRUCTION MANUAL 1WMPD4003356 ...
Page 2: ......
Page 4: ......
Page 30: ......
Page 46: ......
Page 68: ......
Page 83: ...9 5 9 3 External Dimensions 9 3 1 Amplifier Unit AS2503 ...
Page 84: ...9 6 9 3 2 Amplifier Unit AS2603 ...
Page 85: ...9 7 9 3 3 Panel Cut Sizes ...
Page 87: ...9 9 9 3 5 Rack mount Case Rack Rail Elongate Hole ...