142
A&D standard format
This is the standard format for sending data to peripheral devices.
Consists of 16 characters (not including the terminator).
The condition of the data is indicated with a 2-character header.
The data is padded with polarity and zeros (filling the data’s higher order’s surplus part with zeros).
If the data is zero, the polarity is positive.
The unit consists of three characters.
S T
,
+ 0 1
.
2 3 4 5 6 7
g
C
R
L
F
Header
Data
Unit
Terminator
S T
When stable
C
R
:
Carriage return,
ASCII 0Dh
U S
When unstable
L
F
:
Line feed,
ASCII 0Ah
Q T
Counting mode when stable
:
Space,
ASCII 20h
O L
When overloaded
In the external key print mode of the AD-8127 multi-functional
compact printer, a received A&D standard format is printed as shown
on the right.
If the gross/tare weight is added in the [Data to be added] screen of the function table, the second
header corresponding to the output data is appended to the header.
S T
,
N
,
+ 0 1
.
2 3 4 5 6 7
g
C
R
L
F
2nd header
Data
Unit
Terminator
N
Net weight
C
R
:
Carriage return,
ASCII 0Dh
G
Gross weight
L
F
:
Line feed,
ASCII 0Ah
T
Tare weight
:
Space,
ASCII 20h
P T
Preset tare weight
Display settings: MENU key
System settings button
Communication button
Data output button
Data to be added button
[Data to be added] screen
Summary of Contents for BA-125DTE
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315 ...
Page 84: ...84 Output example for periodic check output results Periodic check output results ...
Page 85: ...85 Periodic check results PDF ...
Page 166: ...166 Unit mm Display unit 138 182 73 Ionizer AX ION 25 162 68 129 ...
Page 170: ...MEMO ...
Page 171: ......