163
Individual Specifications
BA-6TE
BA-6DTE
Capacity
6.2 g
6.2 g
2.1 g
Maximum display
6.200084 g
6.20008 g
2.100009 g
Readability
0.001 mg
0.01 mg
0.001 mg
Repeatability
Standard deviation
(measuring point)
0.0010 mg (1 g)
0.0030 mg (6 g)
0.01 mg
(6 g)
0.0025 mg (1 g)
Linearity
±0.010 mg
±0.02 mg
±0.010 mg
Stabilization time
FAST
setting, good
environment
Approx. 10 seconds
Approx. 10 seconds
Approx.10 seconds
Counting mode
Minimum unit weight
0.1 mg
Percent mode
Minimum 100%
reference mass
10.0 mg
Applicable weights for
calibration test/
sensitivity adjustment
Any weight between 1 and 5 g
(factory setting: 5 g)