32
Smart range function
For BA-6DTE / BA-225DTE / BA-125DTE, there are two types of ranges: the standard range and
precision range (high resolution).
Smart range function
The range can switch automatically between the standard range and precision range (high resolution),
depending on the value displayed.
Pressing the RE-ZERO button
to set the displayed value to zero allows for weighing in the
precision range, regardless of the tare value. The range can be fixed to the standard range by using
[Minimum display digit] in the [Application] settings screen.
Display settings: HOME key
APP button
[Application] settings screen
Minimum display digit button Select from [No blank], [1 digit blank], [2 digit blank].
Operation example
Example weighing in g for BA-6DTE
Step 1. Start weighing in the precision range.
Press the RE-ZERO button
to set
the display to zero and enable the
precision range.
Step 2. Place a container on the weighing pan.
When the displayed value exceeds the
precision range, the balance automatically
switches to the standard range.
Step 3. Press the RE-ZERO button
to set
the display to zero and enable the
precision range.
Step 4. Place a sample in the container.
If the weighing value does not exceed
the precision range, the sample can be
weighed in the precision range.
Step 5. Remove the sample and container from
the weighing pan.
Weighing pan
RE-ZERO button
Zero display
Precision range
Container (tare)
Standard range
RE-ZERO button
Zero display
Precision range
Sample
Precision range
Nothing on the pan
Precision range and standard range
Model
Unit
Precision range after pressing the
RE-ZERO button
Standard range
BA-6DTE
g gram
0.000000 g
-
2.100009 g
2.10001 g
-
6.20008 g
mg milligram 0.000 mg
- 2100.009 mg
2100.01 mg -
6200.08 mg
BA-225DTE
g gram
0.00000 g
-
51.00009 g
51.0001 g -
220.0008 g
mg milligram 0.00 mg
-
51000.09 mg
51000.1 mg - 220000.8 mg
BA-125DTE
g gram
0.00000 g
- 51.00009 g
51.0001 g -
120.0008 g
mg milligram 0.00 mg
-
51000.09 mg
51000.1 mg - 120000.8 mg
Summary of Contents for BA-125DTE
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315 ...
Page 84: ...84 Output example for periodic check output results Periodic check output results ...
Page 85: ...85 Periodic check results PDF ...
Page 166: ...166 Unit mm Display unit 138 182 73 Ionizer AX ION 25 162 68 129 ...
Page 170: ...MEMO ...
Page 171: ......