39
Model
100% reference mass
Decimal separator
position
BA-6TE / BA-6DTE
0.0010 g - 0.0099 g
1 %
0.0100 g - 0.0999 g
0.1 %
0.1000 g -
0.01 %
BA-225TE / BA-225DTE / BA-125DTE
0.0100 g - 0.0999 g
1 %
0.1000 g - 0.9999 g
0.1 %
1.0000 g -
0.01 %
Minimum weight alert function
If you configure this display, the HOME screen changes to the weighing screen with the minimum
weight alert function.
Display settings: HOME key
APP button
[Application] settings screen
Application selection button Select [Minimum weight alert function].
HOME key
Weighing screen with the minimum weight alert function
1
Minimum weight alert function
3
Minimum weight alert display
2
Minimum weight setting button
Name
Description
1
Minimum weight alert
function
Displayed when the minimum weight alert function is enabled.
2
Minimum weight setting
button
Displays the [Minimum weight setting] screen.
Also displays the currently stored minimum weight.
3
Minimum weight alert
display
Displayed when the sample weight is less than the set minimum weight.
This is the screen for the minimum weight alert function. This function can be used only in "mg" unit.
Minimum weight alert function usage
Minimum weight is the minimum sample weight required to perform correct quantitative analysis taking
the measurement error of the balance used into account. If the sample amount is too small, the
proportion of measurement error in the measured value increases, and the reliability of the analysis
result thus may drop.
The minimum weight alert function makes it possible to judge immediately whether the sample amount
meets the set minimum weight.
Summary of Contents for BA-125DTE
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315 ...
Page 84: ...84 Output example for periodic check output results Periodic check output results ...
Page 85: ...85 Periodic check results PDF ...
Page 166: ...166 Unit mm Display unit 138 182 73 Ionizer AX ION 25 162 68 129 ...
Page 170: ...MEMO ...
Page 171: ......