42
Automatic (internal weight)
For this setting, the minimum weight is automatically measured with the internal weight in the
[Repeatability measurement mode] screen. When the measurement is complete, the [Minimum weight
measurement result] screen described in “
5.5.4. Minimum weight measurement result
” is displayed.
1
Back button
Name
Description
1
Back button
Returns to the previous screen.
Manual (external weight)
For this setting, measurement is performed with your weight to compute the minimum weight.
The procedure is displayed in the message field.
Weighing example for BA-6DTE
Step 1. Press the RE-ZERO button
.
Step 2. Place a weight on the weighing pan.
Step 3. Remove the weight when the stabilization
indicator is displayed.
Step 4. Repeat steps 1, 2 and 3 ten times.
Step 5. When the measurement is complete, the
[Minimum weight measurement result] screen
described in “
” is displayed.
Weighing pan
RE-ZERO button
Place the weight
Remove
the weight
10 times complete
[Minimum weight measurement result] screen
Summary of Contents for BA-125DTE
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315 ...
Page 84: ...84 Output example for periodic check output results Periodic check output results ...
Page 85: ...85 Periodic check results PDF ...
Page 166: ...166 Unit mm Display unit 138 182 73 Ionizer AX ION 25 162 68 129 ...
Page 170: ...MEMO ...
Page 171: ......