Automatic sensitivity adjustment
Display settings: MENU key
Sensitivity adjustment button
Automatic sensitivity adjustment button
[Automatic sensitivity adjustment] settings
screen
1
2
3
Name
Setting value (setting range)
Description
1
Automatic sensitivity
adjustment execution
condition setting field
OFF, Temperature change ,
Preset time 1,
Preset time 1&2,
Interval
Set the automatic sensitivity adjustment
execution condition.
If you select the preset time or interval, the
relevant time setting is displayed below the
setting field.
2
Automatic sensitivity
adjustment time setting
0:00 - 24:00
Set the automatic sensitivity adjustment
start time.
3
Automatic sensitivity
adjustment interval
setting
0.5 ,
0
1.0,
0
1.5,
0
2.0,
0
2.5,
0
3.0,
0
3.5,
0
4.0,
0
4.5,
0
5.0,
0
5.5,
0
6.0,
0
7.0,
0
8.0,
0
9.0,
10.0, 11.0, 12.0, 14.0, 16.0,
18.0, 20.0, 22.0, 24.0
Set the automatic sensitivity adjustment
interval.
Settings in the red box are default values (factory settings).
This function automatically adjusts the sensitivity of the balance according to ambient temperature
change, set time or interval time using the internal weight. It works even when the display is off. If the
GLP output is enabled, the sensitivity adjustment execution record is output after the sensitivity
adjustment.
The execution condition for the automatic sensitivity adjustment can be based on the temperature
change, preset time or interval.
For preset time, you can configure two settings: “Preset time 1” and “Preset time 2”.
For the interval, you can set a value between 0.5h and 24.0h.
Summary of Contents for BA-125DTE
Page 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315 ...
Page 84: ...84 Output example for periodic check output results Periodic check output results ...
Page 85: ...85 Periodic check results PDF ...
Page 166: ...166 Unit mm Display unit 138 182 73 Ionizer AX ION 25 162 68 129 ...
Page 170: ...MEMO ...
Page 171: ......