43
Minimum weight measurement result
The [Minimum weight measurement result] screen is displayed when the
1
Measurement method
2
Measurement tolerance
3
Repeatability (standard deviation)
4
Minimum weight
5
PRINT button
6
REGISTER button
Name
Description
1
Measurement method Displays the method used for the minimum weight measurement.
2
Measurement tolerance Displays the minimum weight measurement tolerance.
3
Repeatability (standard
deviation)
Displays the standard deviation for this measurement.
4
Minimum weight
Displays the minimum weight.
5
PRINT button
Outputs the minimum weight measurement result to the device connected to
the balance.
6
Register button
Registers the minimum weight.
*
If the repeatability (standard deviation) is 0.41d or less, the minimum weight is 2000 times the 0.41d in
compliance with USP.
"d" is a unit of readability.