9.2.
Initializing the Balance
This function restores the following parameters to the factory settings.
Calibration
data
Function table (excluding the time and date)
Sample unit mass value (counting mode),
100% reference mass value (percent mode)
The data that is stored in the balance using the data memory function
External weight value
Function switch settings ("
9.1. Permit or Inhibit
")
Liquid density and temperature in the density mode
Note Be sure to calibrate the balance after initialization.
1 Press
the
ON:OFF
key to turn the display off.
2 While pressing and holding the
key and the
RANGE
key, press the
ON:OFF
key to display
p5
.
3 Press
the
RANGE
key to display
Clr
.
4 Press
the
key to display
Clr
no
.
To cancel the operation, press the
CAL
key.
5 Press
the
RE-ZERO
key to display
Clr
go
.
6 Press
the
key to initialize the balance.
The balance will automatically return to the weighing mode.