28
Class Item Param-
eter
Description
func
Hold
Display hold function
0
Disabled
d = readability (scale interval or
“division”)
Hold: Holds stable weight values
except in the zero band.
When the value changes more than
the limit, the display hold is canceled.
1
Hold. Limit : ±10d
2
Hold. Limit : ±50d
3
Hold. Limit : ±100d
4
Hold. Limit : ±200d
totl
Accumulation function
0
Disabled
1
Enabled
ltUp
LCD backlight control
0
Always off
To control how the LCD
backlight turns off after the
STABLE indicator turns on.
Weight changes or key
operation will turn the
backlight on.
1
Turns off after 5 seconds
2
Turns off after 10 seconds
3
Turns off after 30 seconds
4
Turns off after 60 seconds
5
Always on
Unit
Weighing units to be
displayed
Sets to display units
Refer to “
8.4. Storing
Weighing Units
”.
id
ID number for GLP output Sets the ID number
Refer to “
10. ID NUMBER
AND GLP
”.
: Factory setting
*
:
When the AD-8127/AD-8121B format is selected, the interval between continuous data
is 1.6 seconds regardless of the setting “
pU5e
”.
**
:
Even if the weight display is “5d”, the sample weight may not be accepted. This is
because the weight display data is rounded off internally.