37
15. SPECIFICATIONS
15-1. Specifications
MODEL EK-300AEP
EK-3000AEP
EK-12KAEP
Intrinsically safe structure
Ex ia op is IIB T3
Ambient temperature range
-25°C to 40°C / -13°F to 104°F
Weighing capacity
300 g
3000 g
12 kg
Readability
0.01 g
0.1 g
1 g
Repeatability
(Std. deviation)
0.01 g
0.1 g
1 g
Linearity
±0.02 g
±0.2 g
±1 g
Sensitivity drift
±20 ppm / °C (10°C to 30°C / 50°F to 86°F)
Number of samples
5, 10, 25, 50 or 100 pieces
Maximum count *
30,000 pcs
30,000 pcs
12,000 pcs
Minimum unit weight *
0.01 g
0.1 g
1 g
% readability
0.1%
Minimum 100 % weight
1 g
10 g
100 g
Display
7 segment LCD with backlight
(Character height 16 mm)
Display update
Approx. 10 times per second
Operating temperature and
humidity range
0°C to 40°C / 32°F to 104°F,
Less than 85% R.H. (non-condensing)
Power source
6.0 VDC, Four “AA”, “LR6” or “AM3” batteries
Operating hours: Approx. 250 hours with backlight off
Weighing pan size
110 mm ø
133 x 170 mm
Weight
(Excluding batteries)
Approx. 1.2 kg
Approx. 1.5 kg
Approx. 1.5 kg
Calibration weight
(factory setting)
300 g
3000 g
10 kg
* In case of “
Umin 0
” (factory setting)
15-2. Option
EJ-12 Carrying
case
15-3. External Dimensions
EK-300AEP
EK-3000AEP / EK-12KAEP Unit: mm
21
8
12
8.
5
71
.5
200
190
110
Φ
54
190
170
12
8.
5
71
.5
200
21
8
13
3
53