9
5. OPERATION
5-1. Turning the power ON and OFF
1. Press the ON/OFF key to turn the power ON.
All the display symbols appear as shown above.
(As for weighing units: Only the units stored in memory are displayed.)
The display turns off except a weighing unit and a decimal point.
The scale waits for the weight value to become stable, and zero is displayed with
the ZERO indicator (power-on zero).
The range for power-on zero is within ±10% of the weighing capacity around the
calibrated zero point.
If the power is turned ON while there is a load beyond this range, the scale is tared
to zero and the NET and ZERO indicators turn on.
2. Press the ON/OFF key to turn the power OFF.
Auto power-off function
The auto power-off function automatically turns the power OFF when zero
is displayed for approximately 5 minutes.
See “
poff
” in “8-5. Function list.”
Display lock function
The display lock function locks the display temporarily to make the weight
value easier to read.
See “
Hold
” in “8-5. Function list.”
LCD backlight function
The LCD backlight turns on when the weight value changes 4d or more (d
= minimum display, so, corresponding to four times as great as the
minimum display) or any key operation is performed. When the weight
value is stable for a certain period of time, the backlight automatically
turns off. The backlight can also be set to remain on or off.
See “
ltUp
” in “8-5. Function list.”