12
5-5. Weighing range
Function setting
Description
rng 0
Fixed range
Fixed to 1/30,000. (R2 range for the EK-30KL)
rng 1
Automatic range function (Only for the EK-30KL)
The EK-30KL has two weighing ranges when the automatic
range function is selected. The indicator
R1
or
R2
turns
on to indicate which range the weight value belongs to.
When the weight value exceeds the maximum value of
the R1 range, the weighing range changes automatically
to the R2 range.
When there is nothing on the weighing pan and zero is
displayed with the ZERO indicator, the weighing range
will automatically be the R1 range.
When the RE-ZERO key is pressed in the R2 range,
the scale is tared to zero and the weighing range changes
to the R1 range.