16
6. COMPARATOR
The results of the comparison are indicated by
HI
,
OK
or
LO
in the display.
The comparison is as follows:
LO
Lower limit value
OK
Upper limit value
HI
Comparison mode (see “
Cp
” in “8-5. Function list”):
No comparison (comparator function disabled).
Compares all data.
Compares all stable data.
Compares plus data except those near zero.
Compares stable plus data except those near zero.
Compares all data except those near zero.
Compares all stable data except those near zero.
“Near zero” is a range within ±4d from the zero point in grams or kilograms.
d = minimum display
The upper limit and lower limit numerical values are common to each of the weighing,
counting and percent modes. An example is shown below.
Upper limit value “001010”: “101.0 g” “1010 pcs” “101.0%”
Lower limit value “000990”: “99.0 g” “990 pcs” “99.0%”
6-1. Setting example
To set “Compares plus data except those near zero.”
Selecting a comparison mode
1. With the power turned OFF, press and hold the
RE-ZERO key and press the ON/OFF key to turn
the power ON and enter the function setting mode.
func
appears.
(If the comparison mode is already set, press the
MODE key to go to “Entering the upper and lower limit
values.”)
2. Press the PRINT key to display
poff
X .
3. Press the MODE key several times to display
Cp
X .
4. Press the RE-ZERO key several times to display
Cp 3
.
5. Press the PRINT key to store the settings.
Cp Hi
appears after
end
.
To store
Press several times
Press several times