7-2. Zero calibration
5. With nothing on the pan, press the RE-ZERO key
to calibrate the zero point. Do not touch the pan
during zero calibration.
When the zero calibration is complete, the calibration
weight value appears in the display.
To perform zero calibration only and finish the
procedure, press the ON/OFF key to turn the
power OFF or press the CAL switch.
7-3. Calibration using a calibration weight
6. Place the calibration weight with the same value as
displayed on the pan. Press the RE-ZERO key to
perform calibration. Do not touch the pan during
calibration.
7.
end
appears.
Remove the weight from the pan.
Press the ON/OFF key to turn the power OFF or
press the RE-ZERO key.
With nothing on the pan
To store
Calibration
weight
To store