31
`~~~~~~~~~~
A
~
&
~
D<CRLF>
MODEL
~~~
EK-15KL<CRLF>
S/N
~~~~
6A6123456<CRLF>
ID
~~~~~~~~
ABCDEF<CRLF>
DATE<CRLF>
<CRLF>
TIME<CRLF>
<CRLF>
CALIBRATED(EXT.)<CRLF>
CAL.WEIGHT<CRLF>
~~~~~
+15000.0
~~
g
<CRLF>
SIGNATURE<CRLF>
<CRLF>
<CRLF>
---------------
~
<CRLF>
<CRLF>
<CRLF>
10-2. GLP report
To print the GLP report to an AD-8121B printer, select the function setting “
info 1
”
and “
pUse 1
” and use MODE 3 of the printer.
To output the GLP report to a personal computer, select the function setting “
info
2
” and “
pUse 0
.”
Calibration report
1. Perform calibration according to “7-3. Calibration
using a calibration weight.”
2.
end
appears when the calibration is
complete.
3.
glp
appears and calibration report is
output.
4.
end
appears again. Remove the weight.
Press the ON/OFF key to turn the power
OFF or press the RE-ZERO key.
AD-8121 format “
info 1
”
General format “
info 2
”
: Space,
ASCII
20h
CR: Carriage return, ASCII 0Dh
LF:
Line feed, ASCII 0Ah
To output the results
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration
executed
Calibration weight
Column for
signature