37
12. SPECIFICATIONS
12-1. Specifications
MODEL
EK-15KL
EK-30KL
Weighing capacity
15 kg
3 kg
30 kg
Minimum display ”d”
0.1 g
0.1 g
1 g
Number of samples
5, 10, 25, 50 or 100 pieces
Maximum count *)
150,000 pieces
300,000 pieces
Minimum unit mass *)
0.1 g
0.1 g
Minimum % display
0.1 %
0.1 %
Repeatability (Standard deviation)
0.3 g
0.3 g
1 g
Linearity
±0.6 g
±0.6 g
±1 g
Sensitivity drift
±20 ppm / °C (10°C to 30°C / 50°F to 86°F)
Display
7 segment LCD with backlight (Character height 22 mm)
Display update
Approximately 10 times per second
Operating temperature range/
humidity
10°C to 40°C / 14°F to 104°F
Less than 85% R.H. (non-condensing)
Power supply
AC adapter or optional lead-acid battery
Battery operating hours
Approximately 100 hours (at 25°C with backlight off)
Weighing pan size
300 mm x 210 mm
Mass (approximately)
4.2 kg
4.9 kg
Calibration weight (factory setting)
15 kg
30 kg
*) In case of “
Umin 0
” (factory setting)
12-2. Other weighing units
MODEL
EK-15KL
EK-30KL
Weighing capacity
529.110
105.820
1058.20
oz.
Minimum display
0.005
0.005
0.05
Weighing capacity
33.0690
6.6140
66.140
lb
Minimum display
0.0005
0.0005
0.005
Weighing capacity
482.260
96.450
964.50
ozt
Minimum display
0.005
0.005
0.05
Weighing capacity
396.830
79.365
793.65
tl
Minimum display
0.005
0.005
0.05
The unit “tl (Hong Kong General / Singapore)” is for special versions only.
12-3. Option
OP-02: Sealed lead-acid battery (Recommended model: YUASA NP4-6)