6
4. SETTING UP
4-1. Setting up your scale
1. Adjust the level of the scale using the leveling
feet. Use the bubble spirit level to confirm.
The bubble should be in the center of the circle.
2. Calibrate the scale before use.
See “7. CALIBRATION.”
Bubble spirit level
Scale location
Observe the following to get the most out of your scale and weigh correctly.
Do not install the scale in locations that are subject to dust, draft, vibration,
excessive temperature changes, condensation, or that may have magnetic
fields.
Install the scale on a solid and level surface.
Do not install the scale in direct sunlight.
Do not install the scale near heaters or air conditioners.
Use a stable AC power source.
Do not install the scale in a place where flammable or corrosive gases present.
Allow the scale to reach equilibrium with the ambient temperature before
use.
Turn the power ON and warm up the scale for at least 30 minutes before use.
When the scale is installed for the first time, or the scale has been moved,
perform calibration as described in “7. CALIBRATION.”
If it seems that the scale is being affected by static electricity, ground the
scale using the grounding terminal to reduce effects of static electricity.
Grounding terminal