background image

6.  Go to step 2 in section “

9-1. Calibration Procedure Using a Weight

” 

to calibrate the remote scale. 

14.

 

SPECIFICATIONS 

 

MODEL FC-500

i

 FC-1000

i

FC-2000

i

FC-5000

i

FC-10K

i

FC-20K

i

 FC-50K

i

Capacity (k)g 

500 g 

1 kg

2 kg

5000 g

10 kg 

20 kg 

50 kg

Resolution (k)g 

0.05 g 

0.1 g

0.2 g

0.5 g

1 g 

2 g 

5 g

Capacity lb 

1 lb 

2 lb

5 lb

10 lb

20 lb 

50 lb 

100 lb

Resolution lb 

0.0001 lb  0.0002 lb 0.0005 lb

0.001 lb

0.002 lb 

0.005 lb 

0.01 lb

Sample Size 

10 pieces normal – 5, 25, 50, 100 or random number, user selectable 

Min. Unit Weight 
(Normal mode) 

0.01 g 

0.02 g 

0.04 g 

0.1 g 

0.2 g 

0.4 g 

1 g 

Min. Unit Weight 
(Fine mode) 

0.0005 g 

0.001 g 

0.002 g 

0.005 g 

0.01 g 

0.02 g 

0.05 g 

Non-linearity 

±0.05 g 

±0.1 g 

±0.2 g 

±0.5 g 

±1 g 

±2 g 

±5 g 

Repeatability 

0.05 g 

0.1 g 

0.2 g 

0.5 g 

1 g 

2 g 

5 g 

Span Drift 

0.002%/°C (5°C~35°C) typ. 

Operating Temp. 

-10°C to 40°C/14°F to 104°F, less than 85%R.H. (No Condensation) 

Display 

7 segment / 5x7 dot VFD 

Display Update 

Approximately 10 times per second 

Interface 

RS-232C (1ch.) standard, Maximum 3 ch. with optional interface 

Power 

AC adapter or Optional Ni-MH Battery Pack 

Battery Operating Time: Approx. 10 hours (main scale only) 

Platform Size 

205 x 152 (mm)   

8.1 x 6.0 (inches) 

315 x 270 (mm) 

12.4 x 10.6 (inches) 

Dimensions  
(W) x (D) x (H) 

330 x 462 x 107 mm 

13.0 x 18.2 x 4.2 inches

330 x 462 x 117 mm 

13.0 x 18.2 x 4.6 inches 

Weight (approx.) 

5.5 kg 

5.5 kg 

6.5 kg 

6.7 kg 

6.7 kg 

6.7 kg 

7.5 kg 

Calibration Weight 

500 g ±   

0.01 g 

1 kg ± 

0.02 g 

2 kg ± 

 0.05 g 

5kg±0.1g 10kg±0.2g 20kg±0.5g  50kg±1g

Accessories 

This manual, AC adapter 

 

 73 

Summary of Contents for FC-5000Si

Page 1: ...Counting Scale FC 50Ki FC 5000Si FC 20Ki FC 500Si FC 10Ki FC 5000i FC 2000i FC 1000i FC 500i WM PD4000541B...

Page 2: ...translated into another language without the prior written consent of the A D Company Product specifications are subject to change without any obligation on the part of the manufacture Compliance with...

Page 3: ...nit Weight By ID Number 18 4 ENTERING A TARE WEIGHT 19 4 1 Using the KEYBOARD TARE Key 19 4 2 To Clear TARE 20 5 STORE UNIT WEIGHT 21 5 1 Store Unit Weight by ID Number 21 5 2 Clearing A Stored Unit W...

Page 4: ...12 1 RS 232C Specifications 46 12 2 Data Output Mode 47 12 3 Connecting the AD 8121 Printer MODE 1 or MODE 2 48 12 4 Connecting the AD 8121 Printer MODE 3 49 12 5 Command Mode 50 12 6 Using a Bar Code...

Page 5: ...eight directly Three UNIT WEIGHT BY LED s will navigate you to store a unit weight easily ACAI Automatic Counting Accuracy Improvement supports counting by recalculating the unit weight when a sample...

Page 6: ...s The counting scale An AC adapter check that the AC input rating is correct Remove the protective packing materials from around the scale and between the pan and scale casing Remove Packing Material...

Page 7: ...111111 9999999 self check 3 Please leave the scale with the AC adapter connected for at least thirty minutes before use We call this situation warm up If desired press the STANDBY OPERATE key to turn...

Page 8: ...near heaters or air conditioners Don t install the scale in direct sunshine Try to ensure a stable AC power supply when using the AC adapter Keep equipment containing magnets away from the scale Warm...

Page 9: ...mode only front panel keys that would be used in 3 3 Unit Weight By Samples counting operations are active All others will not operate The following keys are active in the simple operation mode Keys...

Page 10: ...llowing 1 When three UNIT WEIGHT BY LED s are blinking at display ON or if the RESET key has been pressed 2 Press the ID key id 000000 will be displayed with 000000 blinking 3 Press the ENTER key The...

Page 11: ...t data is being accumulated The REMOTE SCALE key switches between the main and a remote scale if used The PRINT key sends count weight or unit weight data The KEYBOARD TARE key allows entering a known...

Page 12: ...zero as long as the weighing pan is empty or within 2 of capacity 1 Remove everything from the weighing pan and press the ZERO key Then the display shows and waits for the weighing data to become stab...

Page 13: ...ble 4 The scale will subtract the weight of the container and the weight display changes to net weight The TARE ENTERED indicator will be displayed STABL 8888810 881 pcs W2WEIGHT kg 12134567890 000 23...

Page 14: ...he section noted for more instructions STABL 8888810 881 pcs W2WEIGHT kg 12134567890 000 23ASAMPUW g 123415670 000004567890123456 7 TARE ENTERED ZERO M SCALE TOTAL UNIT WEIGHT BY KEY BOARD You can ret...

Page 15: ...then add the additional number of sample pieces displayed You can ignore the Add Sample message and continue counting by pressing the ENTER key However the results may not be accurate Refer to F Funct...

Page 16: ...key The display will show for a moment while calculating the unit weight After a moment the display will show the count total weight and unit weight If the Add Sample display appears again then the s...

Page 17: ...ill show for a moment while calculating the unit weight After a moment the display will show the count total weight and unit weight If the Add Sample display appears again then the sample size is not...

Page 18: ...culating the unit weight After a moment the display will show the count total weight and unit weight If the Add Sample displays again then the sample size is not enough for accurate counting add the a...

Page 19: ...t 32g 4 Press the ENTER key The unit weight 32g will have been entered If the unit weight entered is too light lo ut low unit weight will be displayed and you will be returned to step 3 5 You may now...

Page 20: ...unit weight entered for the ID number you tried to recall no id will be displayed and you will be returned to step 3 5 You may now begin counting operations for pieces of the same weight 0 881 pcs W2...

Page 21: ...Use the 0 9 and 10 key pad to display the desired TARE weight If you hit the wrong key press the C key to clear and start again Example of a tare weight 615g 4 Press the ENTER key The weight display...

Page 22: ...ndicator will be turned off STABL 888 554 881 pcs 2AAWEIGHT kg 1234105678 0 615 23SAaaMPUW g 1234151671 11000456789012345 67 TARE ZERO M SCALE TOTAL ENTERED STABL 888 510 881 pcs A2WEIGHT kg 121345678...

Page 23: ...ng key press the C key to clear and start again 4 Press the ENTER key The ID number is stored and the display returns to normal If the same ID number was previously stored the scale beeps twice and th...

Page 24: ...all will appear with all blinking 2 Press the ENTER key then all blinking will stop 3 Press the ENTER key again to clear all of ID memories Press the RESET to exit without clearing ID memories The dis...

Page 25: ...the 3 DEF key to place D A A D 7 Press the TOTAL key twice to shift the cursor A A D _ 8 Press the 2 ABC key several times to place C A A D C 9 Press the M A a key to change the symbol A to a a A D C_...

Page 26: ..._ 1 A B C a b c 2 D E F d e f 3 G H I g h i 4 J K L j k l 5 M N O m n o 6 P Q R S p q r s 7 T U V t u v 8 W X Y Z w x y z 9 0 Clear Space STORE UNIT WEIGHT ID number input Item code input PRINT To shi...

Page 27: ...e weight by any method If necessary set the comparator limits and use the M accumulation 2 Go to step 2 of section 5 1 Store Unit Weight By ID Numbers When you recall a unit weight by the ID key the t...

Page 28: ...2 Press the M key every time you want to add to the count Remember that you may only add the count data once the scale must return to near zero before it will let you add again To Erase the Last M Add...

Page 29: ...g the M Total 1 Press the TOTAL key The count display will show the total count and the TOTAL indicator will come ON The number of additions to the M memory is also shown 2 Press the TOTAL key again T...

Page 30: ...limits Count Weight data GO Lower limits Count Weight data Upper limits LO Count Weight data Lower limits 888 100 881 pcs 2AAWEIGHT kg ABDEHOKLOaa 121341567811 100578id 123456 2A3SAAMUUW g 2BCDEFGHIJ...

Page 31: ...0 to use this mode 1 In step 1 above press the ENTER key the upper limit value will blink 2 Change the upper limit using the 10 key pad and press the ENTER key The upper limit will stop blinking and...

Page 32: ...rom 00 seconds 5 Press the STANDBY OPERATE key to exit and press it again to turn the display back ON Using the Key to Set the Time and Date Set the F Function f 09 01 0 to use the key as this functio...

Page 33: ...30 minutes before starting calibration 9 1 Calibration Procedure Using a Weight The scale should have power connected at least one half hour to warm it up before starting the calibration procedure 1...

Page 34: ...t should be an error will be displayed Check if the weight is correct and try again 6 Press the CAL switch and re attach the calibration switch cover End of the calibration procedure ENTER 8Call00 881...

Page 35: ...ed in the scale 2 Use the 0 9 and 10 key pad to display the desired gravity acceleration value Example of the value 9 800 m s2 3 Press the ENTER key The scale will store the new value If necessary to...

Page 36: ...clears the input setting if you press the wrong key and want to re enter If you make a mistake and want to escape without saving any changes made after the last time the ENTER key was pressed press t...

Page 37: ...tem 7 When finished Press the STANDBY OPERATE key to exit Then press it to turn the display back on The new settings will operate as set ENTER f 05 02 881 pcs Comparator 12345678C000045678id 000000 23...

Page 38: ...ieces Or disable the Add sample request function 0 Add sample request function is disabled Light unit weight can be accepted without Add more sample request 13 The unit weight can be entered without r...

Page 39: ...0003 ID memory contains the unit weight and item code only You select which data to be stored by keying in a 0 or 1 for the data tare weight comparator limits or total count Example Key in 0 1 0 to d...

Page 40: ...m Unit Weight 1d 1 weighing display division The factory setting may be different for some countries 0 1 5 d FC i series 1 20 d FC Si series 13 1 100 d FC i series 1 400 d FC Si series F 03 X M Accumu...

Page 41: ...detection poor environment Beeper for key operation 03 Beeper ON 1 Beeper OFF F 05 X Comparator Comparator Mode 03 comparator off 1 compare all data 2 compare stable data 3 compare all data except wh...

Page 42: ...07 X F 08 X Data Output f 06 X for standard RS 232C f 07 X for Ch 1 PORT 1 and f 08 X for Ch 2 PORT 2 Data Out Mode 03 Key mode Data is sent by pressing the PRINT key command mode 1 Stream mode Data...

Page 43: ...and the count f X 02 06 07 08 00000 Weight PCS count Date Time Unit ID Weight Data Format 03 Format for AD 8121 MODE 1 1 Format for AD 8121 MODE 3 No difference between 0 and 1 when used with the UFC...

Page 44: ...data format WT 001 2346 kg 3 Decimal point Stable weight data format WT 001 2346 kg F 10 X External Control Signal Input Control Signal Input 1 and 2 Select an input key function as listed below for...

Page 45: ...F 11 X Time Date Time and Date Display Select the order of date display 0 Year Month Date 13 Month Date Year 2 Date Month Year f 11 01 43...

Page 46: ...ounting results for every different batch of the same items use ACAI every time you start counting the next batch The ACAI does not work when you use a remote scale if the unit weight was set on the m...

Page 47: ...calculate the unit weight until the ENTER key is pressed as long as it is at the proper time and the guidelines in the ACAI notes have been followed The ACAI manual mode is controlled by F Function f...

Page 48: ...x Data format Baud rate 2400 4800 9600 bps Data 7 bits parity 1bit even or odd or 8 bits non parity Start bit 1 bit Stop bit 1 bit Code ASCII Terminator Data Send CRLF CR 0Dh LF 0Ah Data Receive CR or...

Page 49: ...06 07 08 01 2 Data is sent if the WEIGHT display is stable at 5d weighing display division and above The next transmission can not occur until after the weight display falls below 5d Auto print Mode...

Page 50: ...y or weight data only To print date and time use the AD 8121 s calendar clock function and set f 06 07 08 2 to print pcs count data only or weight data only Print Operations Settings Print By F Functi...

Page 51: ...otal data accumulated by the M key will be printed along with the number of additions to M memory The AD 8121 MODE 3 does not have statistical functions AD 8121 Printout Sample MODE 1 MODE 3 ID number...

Page 52: ...from ID memory F123CRLF recalls from ID000123 G Set a known unit weight G 0 123CRLF sets the unit weight as 0 123g or 0 123 lb I Same as the REMOTE SCALE key J Same as the TOTAL key Key command K Sam...

Page 53: ...command requests a data reply the scale will send the data For other commands the scale will send an acknowledgment AK 06H upon acceptance of the command If the command is I S T or Z the scale will s...

Page 54: ...eply Q T 0 0 0 0 1 2 3 4 P C C LF Stable positive data U S 0 0 0 0 5 6 7 8 P C C LF Unstable negative data O L 9 9 9 9 9 9 9 9 P C C LF E display Weight data Command W T C LF Reply S T 0 0 1 2 3 4 6 k...

Page 55: ...total count 9 digits including sign number of additions 8 digit CR LF Command M R C LF Request the contents of the next ID Reply M R 0 0 0 1 2 4 A B C D E F G H I J K L 2 3 4 5 6 7 8 0 0 0 2 3 4 5 0 0...

Page 56: ...ion Maximum 8 digits Request the scale to reply with the F Function settings Command F C 0 5 0 1 C LF Request the setting of f 05 01 Reply F C 0 5 0 1 0 C LF f 05 01 0 Command F C C LF Request to repl...

Page 57: ...In most cases a bar code reader with an RS 232C interface is to connect directly with a personal computer The signals must be converted to connect this kind of bar code reader with the scale Refer to...

Page 58: ...Identification Code Numbers Function ID Number 01 6 digits ID number Recall the unit weight by ID number Similar function to 3 5 Unit Weight By ID Number Unit Weight g 12 10 digits including sign and...

Page 59: ...valid identification code The data is out of range The tare data is more than the capacity etc The tare or unit weight has a minus sign The data has start stop characters or a check digit These will l...

Page 60: ...Comparator upper limit 12345 Codabar NW 7 A17 000012345A 58...

Page 61: ...if the other format data is not necessary to send Store Text Data into the Scale Memory Command P F P C T E X T 2 0 S P 2 C R L F W T C R L F C LF Terminator Reply AK The PF command sends text data th...

Page 62: ...ers for the scale data will be replaced by the format below when the scale sends them out Data has a fixed number of digits including a sign and a decimal point The insignificant zeros are replaced by...

Page 63: ...LF Terminator codes Examples of PF command and AD 8121 Printout Sample AD 8121 Printout PF Command f 06 07 08 03 0 or 1 ID 000123 shows Space Normally the printer needs to receive the terminator and d...

Page 64: ...ORT 1 set f 07 01 5 Refer to 12 6 Using a Bar Code Reader about the bar code data and prepare bar code labels The OP 01 bar code reader can be connected to one of the RS 232C interfaces It can read ba...

Page 65: ...odes step by step then the bar code reader will be set to work with the scale This sets the RS 232C to be 9600 bps and 8 bits data non parity Set the F Function f 06 07 08 04 2 and f 06 07 08 05 2 For...

Page 66: ...tting EAN 13 check digit Step 8 Disable transmitting EAN 8 check digit Step 9 Disable transmitting CODABAR Start Stop character Step 10 Disable transmitting CODE 39 check digit Step 11 Disable transmi...

Page 67: ...Lock Screw OP 02 Slot 2 Tighten the lock screw on the battery pack 3 When desired turn the battery switch on to supply power to the scale The display will come on after its self check 4 Use the STANDB...

Page 68: ...e sure to charge the battery pack when using for the first time or if it was not used for long time more than one month Two or three times of recharging may be needed to reach full charge Be sure to u...

Page 69: ...C 1 Disconnect the AC adapter from the scale 2 Remove the two screws and panel covering the OP 03 04 slot 3 Connect the connector in the scale to the OP 03 unit OP 03 04 slot 4 Secure the OP 03 unit u...

Page 70: ...me as the standard RS 232C interface Refer to 12 RS 232C SERIAL INTERFACE Comparator Relay Output RS 232C Ch 1 Comparator Relay Output Circuit Maximum rating of the Relay Output The maximum rating of...

Page 71: ...tive screw cap to the remote scale connector Load cell Notes on Using the Remote Scale To get a highly accurate counting unit weight use the main scale to register the unit weight and use ACAI feature...

Page 72: ...is follows Example Load Cell Capacity 100kg A Rated Output 3mV V B Min Division of Display 0 01kg D 5 000 x x B D A X V When a single Load Cell is used the following formula should apply System desig...

Page 73: ...0 1 dividing Ns by 10 In example 1 the new d 5 1 No d Ns 5 20 000 1 4 000 200kg x 0 05kg In example 2 the new d 1 and Ns 3 000 1 No d Ns 1 3 000 30kg x 0 01kg Using the new 1 No go to step 6 and repea...

Page 74: ...capacity Example of capacity 2000kg 4 Press the ENTER key The capacity value stops blinking When the remote scale has been changed or to change capacity go back to step 3 5 Press the ENTER key again...

Page 75: ...g Non linearity 0 05 g 0 1 g 0 2 g 0 5 g 1 g 2 g 5 g Repeatability 0 05 g 0 1 g 0 2 g 0 5 g 1 g 2 g 5 g Span Drift 0 002 C 5 C 35 C typ Operating Temp 10 C to 40 C 14 F to 104 F less than 85 R H No Co...

Page 76: ...tely 10 times per second Interface RS 232C 1ch standard Maximum 3 ch with optional interface Power AC adapter or Optional Ni MH Battery Pack Battery Operating Time Approx 10 hours main scale only Plat...

Page 77: ...ons 280 300 282 117 FC 2000i FC 5000i FC 10Ki FC 20Ki FC 50Ki 285 330 152 462 437 282 205 107 FC 500i FC 1000i 285 330 462 437 165 282 165 106 FC 5000Si 285 330 128 FC 500Si 462 437 282 128 106 285 33...

Page 78: ...s2 Cape Town 9 796 m s2 Rio de Janeiro 9 788 m s2 Chicago 9 803 m s2 Rome 9 803 m s2 Copenhagen 9 815 m s2 San Francisco 9 800 m s2 Cyprus 9 797 m s2 Singapore 9 781 m s2 Djakarta 9 781 m s2 Stockholm...

Page 79: ...World map 77...

Page 80: ...2 73 Capital letter 23 24 Comparator function 28 Comparator limit 25 Comparator result 40 Lower limit 28 29 40 Upper limit 28 29 40 Count data 26 Counting 12 Data format 41 46 48 52 Data length 41 Dat...

Page 81: ...input 42 Slot OP 02 5 64 OP 03 04 5 66 Remote scale 68 Span calibration 32 Spirit level 5 Stable detection speed 39 STABLE indicator 9 Standby indicator 7 10 STANDBY OPERATE key 5 7 9 64 Start bit 46...

Page 82: ...MEMO 80...

Page 83: ...MEMO 81...

Page 84: ...key Space code Standby indicator 82...

Page 85: ......

Page 86: ......

Reviews: