7. COMPARATOR 

 

The results of the comparison are indicated by HI, OK or LO on the display. 
The formula to compare is as follows: 
 

LO < Lower limit value = OK = Upper limit value < HI 

 

Operating conditions (see the “Function list F6”): 

F6-0: No comparison (comparator function disabled). 
F6-1: To compare all data. 
F6-2: To compare more than +4d or less than -4d. 
F6-3: To compare all stable data. 
F6-4: To compare stable data more than +4d or less than -4d. 
F6-5: To compare stable data more than +4d. 
d = minimum weight display (see “12-1 Specifications”) 
In case of counting mode, “d” is equal to minimum weight display of kg mode. 

?

 The upper limit and lower limit numerical values are common to each of the 

weighing and counting mode. 

?

 Ignore the decimal point of setting value to apply it to each mode. 

Example of FG30K / setting value is “001000”. 

Display mode 

Limit value 

Display capacity 

NORMAL resolution kg 

10.00 kg 

30.00 kg x 0.01 kg 

HIGH resolution kg 

1.000 kg 

30.000 kg x 0.005 kg 

HIGHER resolution kg 

1.000 kg 

30.000 kg x 0.002 kg 

NORMAL resolution lb 

10.00 lb 

60.00 kg x 0.02 lb 

HIGH resolution lb 

10.00 lb 

60.00 kg x 0.01 lb 

HIGHER resolution lb 

1.000 lb 

60.000 kg x 0.005 lb 

Counting mode 

1000 pcs 

 

 

Entering the upper and lower limit values 

1. Press  the    MODE   switch  to select weighing 

unit ”kg”, ”lb” or ”oz” 

2. Press and hold  the    MODE   switch to enter  the 

upper limit setting mode. 

3. Enter an upper limit value using the following switches. 

  RE-ZERO   or    ZERO      To select the digit blinking to change. 

  PRINT     

 

 

Increment the value of the selected digit. 

The minus sign can be set at the next digit of the least 
significant digit. The    PRINT   switch alternates the 
minus sign on and off. The blinking  “-” shows minus 
and no sign is plus. 

4. After setting all of digits, press the    MODE   switch. Then, the upper limit is stored 

and the display goes to the lower setting mode. 

5. Set the lower limit in a similar way, and press the    MODE   switch to return to 

weighing mode. 

?

 The upper and lower limits are maintained even if the scale is powered off. 

  

  

0.00 

 0000.00 

HI 

-0000.00 

HI 

Summary of Contents for FG-60KAL

Page 1: ...F FG G S Se er ri ie es s D Di ig gi it ta al l P Pl la at tf fo or rm m S Sc ca al le es s FG 60KAL FG 150KAL FG 30KAM FG 60KAM FG 150KAM FG 30KBM FG 60KBM FG 150KBM WM PD4000966...

Page 2: ...h if not avoided could result in death or serious injury CAUTION A potentially hazardous situation which if not avoided may result in minor or moderate injury This is a hazard alert mark This mark inf...

Page 3: ...on 9 6 COUNTING MODE 10 7 COMPARATOR 11 8 CALIBRATION 12 8 1 Calibration using a weight 12 8 2 Gravity acceleration correction 13 9 FUNCTIONS 14 9 1 The procedure for setting parameters 14 9 2 Functio...

Page 4: ...ation As power source you can use an AC adapter or C size dry batteries The counting function easily counts the number of articles of the same weight The comparator function compares the display value...

Page 5: ...HI OK LO indicator PRINT switch ON OFF switch MODE switch RE ZERO switch Spirit level FG KBM series Display column Column support foot FG KAM only Spirit level Weighing pan Display pod Leveling foot F...

Page 6: ...pcs for U S A models Switches PRINT Switch Outputs the weight value to printer In the setting mode this switch is used to increment the value of the selected digit blinking ON OFF Switch Used to turn...

Page 7: ...turn on The TARE switch will tare the scale when the weight is plus value In this case the ZERO and NET indicators turn on The ZERO operation clears tare operation previously done and NET indicator t...

Page 8: ...column feeding all of slack of the cable on the base into the display column Take care not scratch the cable 6 Put the bundle of cable back into the display column 7 Make sure the cable is fitted to t...

Page 9: ...power source To use the AC adapter insert the AC adapter plug into the AC adapter jack on the rear side of display pod Confirm the AC adapter type is correct for your local voltage When using the bat...

Page 10: ...the power will be switched OFF Auto power off function It is possible to have the power automatically switched OFF if zero is displayed for approximately 5 minutes See 9 2 Function list and set the f...

Page 11: ...ring a value Do not press switches with a sharp implement such as a pencil Do not apply a shock load to the scale Do not place a load onto the pan that exceeds the capacity Keep the scale free from fo...

Page 12: ...pan or in the container 6 Press the MODE switch to calculate and store the unit weight Remove the sample The scale is set to count objects with this unit weight The total weight of sample pieces shoul...

Page 13: ...esolution kg 1 000 kg 30 000 kg x 0 005 kg HIGHER resolution kg 1 000 kg 30 000 kg x 0 002 kg NORMAL resolution lb 10 00 lb 60 00 kg x 0 02 lb HIGH resolution lb 10 00 lb 60 00 kg x 0 01 lb HIGHER res...

Page 14: ...lace something on the pan to disable the auto power off function 2 Press and hold the calibration CAL switch until Cal 0 appears and release the switch The weighing unit must be kg or lb to enter cali...

Page 15: ...he scale Change the gravity acceleration value of the scale to the value of area where the scale will be used Refer to the gravity acceleration map appended to the end of this manual Note It is not ne...

Page 16: ...u do not need to change the parameter go to next step without setting 4 Press the MODE switch Then the display goes to the next function item In this stage the new parameter is not stored in the scale...

Page 17: ...only f 4 0 2400 bps f 4 1 4800 bps RS 232C Baud rate f 4 2 9600 bps f 5 0 Stream mode f 5 1 Command mode f 5 2 Print switch mode f 5 3 Auto print mode data RS 232C Data output mode f 5 4 Auto print mo...

Page 18: ...lation prepare an interface cable using attached DIN connector Or there is a way to use the optional RS 232C cable see 10 2 OP 23 RS 232C serial interface 1 Disconnect the AC adapter from the scale 2...

Page 19: ...nd data output mode F4 F5 and F8 A DIN 8 pin connector JA TCP0586 is provided with the OP 23 for wiring There are optional cables to connect with a personal computer AX KO577A 200 FG to D Sub 25 pin c...

Page 20: ...There are 4 types of units _ k g Weighing data gram _ P C Counting data pcs _ l b Weighing data decimal pound _ o z Weighing data decimal ounce The terminator is always CRLF Example of output data Wei...

Page 21: ...the display returns to between 4d and 4d Auto print mode data Function setting F5 4 The scale transmits the weighing data when the display is stable STABLE indicator is on and the data is more than 4...

Page 22: ...p Reply CR LF When F8 1 is selected there is no reply except the Q command 10 3 OP 24 RS 232C serial interface and Comparator relay output The OP 24 has an RS 232C series interface and relay output fo...

Page 23: ...that the weight sensor itself may have a failure Unit weight error The sample weight is too light to set the unit weight in the counting mode Increase the sample numbers Low battery Warning to show t...

Page 24: ...Std deviation 0 005 kg 0 01 kg 0 02 kg 0 01 kg 0 02 kg Linearity error 0 01 kg 0 02 kg 0 05 kg 0 02 kg 0 05 kg Sensitivity drift 20 ppm C 10 C 30 C 50 F 86 F Display 7 segment LCD display Character h...

Page 25: ...23 12 2 Dimensions FG 60KAL FG 150KAL FG 30KAM FG 60KAM FG 150KAM FG 30KBM FG 60KBM FG 150KBM 530 390 289 300 440 99 128 781 380 220 82 300 300 202 118 781 89 230 380 300 300 220 145 89 118 mm...

Page 26: ...m s2 Cape Town 9 796 m s2 Rio de Janeiro 9 788 m s2 Chicago 9 803 m s2 Rome 9 803 m s2 Copenhagen 9 815 m s2 San Francisco 9 800 m s2 Cyprus 9 797 m s2 Singapore 9 781 m s2 Djakarta 9 781 m s2 Stockho...

Page 27: ...25 World map...

Page 28: ...26 MEMO...

Page 29: ...27 MEMO...

Page 30: ...28 MEMO...

Page 31: ...ngdon Business Park Abingdon Oxon OX14 1DY United Kingdom Telephone 44 1235 550420 Fax 44 1235 550485 German Sales Office Gro e Stra e 13 b 22926 Ahrensburg GERMANY Telephone 49 0 4102 459230 Fax 49 0...

Reviews: