4
EN
1. Introduction
This quick start guide describes the basic functions and operations of the balance.
The functions and operating procedures of the scale (FG-CWP series) are explained in details in
the following instruction manual.
[FG-CWP Series Instruction Manual]
This is an instruction manual for understanding and fully utilizing the water proof digital platform
scale FG-CWP series. This manual can be downloaded from our website
2. Cautions
2.1. Precautions for Installing the Scale
□
Ground the scale, so that the user will not be subjected to an electric shock.
□
To prevent electric shock, do not handle the USB cable with wet hands.
□
The USB cable is not waterproof.
Install the scale in an area where it will not get wet.
□
Do not install the scale where flammable or corrosive gas is present.
□
Do not put excessive force on cables.
□
The scale is heavy. Use caution, as lifting may cause it to fall over.
Consider the following installation conditions to get the greatest performance
from your scale.
□
The ideal conditions for installation are stable temperature and humidity, solid and level
surface, location with no draft or vibration, indoors out of direct sunlight and a stable power
supply.
□
Do not install on a soft floor or where there is vibration.
□
Do not install where there is wind or large fluctuations in temperature.
□
Avoid locations in direct sunlight.
□
Do not install in locations with strong magnetic fields or strong radio signals.
□
Do not install where static electricity is likely to occur. When humidity is 45% R.H. or less,
plastic and insulating materials are apt to be charged with static electricity due to friction, etc.
□
Unstable AC power supply may cause malfunctions.
□
Remove the protective film from the weighing pan before use.
□
Sensitivity adjustment should be performed for more accurate weighing when the scale is first
installed or when it is moved far. Refer to the [Instruction Manual: FG-CWP Series] on our
website https://www.aandd.jp.
2.2. Precautions for Storing
□
Do not disassemble the scale.
□
If the scale becomes dirty, wipe it with a soft cloth slightly moistened with a mild detergent.
Do not use organic solvents.
□
Do not scrub with a brush or the like.
□
Do not spray with strong jets of water.
DANGER
Summary of Contents for FG-CWP Series
Page 44: ...44 CT 2 2 2 3 FG CWP https www aandd jp ZERO 3 FG 30KCMWP FG 60KCLWP FG 150KCLWP AC...
Page 46: ...46 CT FG CWP https www aandd jp 4 2 ON OFF SAMPLE pcs PRINT MODE ZERO 2 ZERO TARE...
Page 47: ...47 CT 5 1 2 3 4 2 1 3...
Page 49: ...49 CT 6 3 error0 error3 error4 error5...
Page 51: ......