- 12 -
The maximum number of text data is 300 characters.
To use the UFC function, the text data has to be sent to the scale from the
computer using “PF” command in the command mode (F06-1) in advance.
Then, connect the scale with the printer and set the F06-2, F06-3 or F06-4.
When the
key is pressed or by auto-print mode (F06-2, F06-3 or F06-4),
the scale will send the stored text data with the parameters replaced by the original
data.
Store Text Data into the Scale Memory
Command P F , $ P
C , ‘ T E X T
’
, # 2 0 , $ S P
½
2 , &
$ C R , $ L F , $ W T , $ C R , $ L F C L
F
Reply P
F C L
F
The “PF” command sends text data that will include:
Parameters for the scale data and control codes.
Parameter
Data & Code
Parameter
Data & Code
$WT Weight
$CM Comma
(2CH)
$TR
Tare weight in use
$SP
Space (20H)
$CP Comparator
result
$CR
C
R
(0DH)
$OK Target
weight
$LF L
F
(0AH)
$HI
HI limit value or upper limit
weight
$LO
LO limit value or lower limit
weight
These parameters must be used in capital letters.
ASCII text string
A text string is described in single quote marks as ‘Data’. The single quote itself is
written as ’’(2 single quotes).
Example:
Text ABC is described as ‘ABC’.
Text ‘ABC’ is described as ‘’’ABC’’’.
The ASCII hexadecimal code
The ASCII hexadecimal codes are written in the form “#" + 2 hexadecimal digits.
This will mainly be used to send control codes that can’t be described as a text string.
Example:
#04
“EOT” of ASCII code
Repeat data
The control codes $SP, $CR and $LF can be used with “ plus a 1 or 2 digit
number”. That code will be repeated the number of times designated.
Example: $LF 9
Repeat “$LF” 9 times.
$SP 12 Use 12 “Spaces”.
Link mark “&”