- 7 -
Out of rage (+)
O L
,
+ 9 9 9 9 9 9 9 9 _ k g
C
R
L
F
Data output mode (F06)
Stream Mode (F06-0)
Data is sent continuously. The data-update rate is approximately 20 times per
second. But for 2400 bps, it is too slow to send 20 times. If necessary, set the baud
rate 4800 bps or more.
Command Mode (F06-1)
The scale is controlled by commands that come from an external device, computer
etc. See “2-3. Command mode” in detail.
Print Key Mode (F06-2)
When the weight display is stable, data is sent by pressing the
key. The
PRINT annunciator will turn on for a short while when the data has been sent.
Auto-print Mode + data (F06-3)
Data is sent if the weight display is stable at +5d (d = weighing display division) and
above. The next transmission can not occur until after the weight display falls below
+5d.
Auto-print Mode +/- data (F06-4)
Data is sent if the weight display is stable at ±5d (d = weighing display division) and
above/below. The next transmission can not occur until after the weight display falls
between –5d and +5d.
Baud Rate (F04)
Select the baud rate according to the device to be connected.
2400 bps (F04-0)
Select 2400 bps to connect with an AD-8121.
4800 bps (F04-1)
9600 bps (F04-2)
2-3. Command Mode
In the command mode, the scale is controlled by commands that come from an
external device, computer etc.
Command List
Command Definition
Notes
Q
Send data immediately.
Z
Zero the scale when the weight is stable.
Same as the ZERO key.
T
Tare the scale when the data is stable.
Same as the TARE key.
U
Switch the weighing unit.
Same as the UNITS key.
D
Switch the analog sweep display mode.
Same as the DISP. key.
PT
Set a known (preset) tare weight.
Set “+” and 6 digit number
without decimal point.