19
5 Confirm that there is nothing on the pan and press the
PRINT key. The balance measures the zero point and
displays the measured value. Do not allow vibration or
drafts to affect the balance.
The balance displays the target weight value.
6 Place the displayed target weight on the pan and press
the PRINT key. The balance measures the target weight
and displays the measured value. Do not allow vibration
or drafts to affect the balance.
7 The balance displays
end
. Remove the weight from the
pan.
8 When the GLP output is set, the balance displays
glp
1
and outputs “Calibration Test Report”.
9 The balance automatically returns to the weighing mode.