10
Press the RE-ZERO key before each weighing to eliminate possible errors.
Calibrate the balance periodically so as to eliminate possible errors.
Take into consideration the affect of air buoyancy on a sample when more accuracy is
required.
Prevent foreign materials, such as powder, liquid and metal, from invading the area around the
weighing pan.
The breeze break (FX-120
i
WP /200
i
WP
/300
i
WP only) is provided as an accessory. An
anti-static treatment has been applied to the breeze break components, but they may be
charged with static electricity when they are unpacked or when the humidity is low. If the
weight value is unstable even when there are no drafts or the balance has a problem with
repeatability, remove the breeze break. Or wipe the clear plates with a moistened cloth, use
the optional AD-1683 DC static eliminator or apply an anti-static spray.
3-3 After Use
Do not allow the balance to be immersed in water. Even though the balance complies with IP
code, the balance will not withstand being completely immersed in water.
Avoid mechanical shock to the balance.
Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or
repair.
Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that
is moistened with warm water and a mild detergent.
The edge of the weighing pan is sharp. Use much care when cleaning the pan.
3-4 Power Supply
When the AC adapter is connected, the balance is in the standby mode if the standby
indicator is on (refer to “3-5 Display Symbols and Key Operation”). This is a normal state and
does not harm the balance. For accurate weighing, warm up the balance for at least 30
minutes before use.