25
5-5 Statistical Calculation Mode (Example of Use)
Here, as an example of use of the statistical calculation mode, mixing of
the multiple formulae such as medicine is described. The mixing
process is recorded using the balance and the printer.
In the example, the FX-3000
i
WP and the AD-8121B (dump print mode)
are connected using the RS-232C serial interface.
5-5-1 Getting Started
Changing the function table
Changes :
To enable the statistical calculation mode
To enable “Zero after output”
Enabling the statistical calculation mode
1 Enter the function table menu.
Press and hold the
SAMPLE
key until
ba5fnc
of the
function table is displayed, then release the key.
2 Select the application function.
Press the
SAMPLE
key several times to display
ap fnc
.
Then, press the
key to display
°
apf 0
.
3 Change the application function parameter to “
2
”.
Press the
RE-ZERO
key to display
apf 2
.
Press the
key to confirm the change.
After
end
,
ba5fnc
is displayed.
Enabling “Zero after output”
4 Select “Zero after output”.
Press the
SAMPLE
key several times to display
dout
.
Then, press the
key to display
°
prt 0
, and press
the SAMPLE key several times to display
°
ar-d 0
.
5 Enable “Zero after output”.
Press the
RE-ZERO
key to display
ar-d 1
.
Then, press the
key to confirm the change. After
end
,
5if
is displayed.
Returning to the weighing mode
6 Press
the
CAL
key to return to the weighing mode.
Return to the weighing mode
Press
several times
Press and hold
Press
several times
Press
several times