30
7-3 Calibration Using an External Weight
This function calibrates the balance using an external weight.
Operation
1 Connect the AC adapter and warm up the balance for
at least 30 minutes with nothing on the pan.
2 Press and hold the
CAL
key until
Calout
is
displayed, and then release the key.
3 The
balance
displays
Cal 0
.
If you want to change the calibration weight (a list of
usable weights is shown on page 28),
press the
SAMPLE
key and proceed to step 4.
If you use the calibration weight value stored in the
balance, proceed to step 5.
4 Specify the calibration weight value as follows:
SAMPLE
key To switch the display condition to:
“All of the segments blinking”
(calibration weight selection mode) or
“The last two digits blinking” (value
adjustment mode).
RE-ZERO
key (To increase the value)
MODE
key
(To decrease the value)
To select the calibration weight or
adjust the value.
key
To store the new weight value. Even
if the AC adapter is removed, the
data is maintained in non-volatile
memory.
CAL
key
To cancel the operation and return to
_
Cal 0
.
Select a
weight
To next page
Press and
hold the key
Release
the key
e.g. Calibration
weight value
1000.12 g