32
7-4 Calibration Test Using an External Weight
This function tests the balance weighing accuracy using an external weight and outputs the result. This
is available only when the “GLP output (
info
)” parameter is set to “
1
” or “
2
”. (Calibration test does not
perform calibration.)
Operation
1 Connect the AC adapter and warm up the balance
for at least 30 minutes with nothing on the pan.
2 Press and hold the
CAL
key until
CCout
is
displayed, and then release the key.
3 The
balance
displays
CC 0
.
If you want to change the target weight (a list of
usable weights is shown on page 28), press the
SAMPLE
key and proceed to step 4.
If you use the target weight value stored in the
balance, proceed to step 5.
4 Specify the target weight value as follows:
SAMPLE
key To switch the display condition to:
“All of the segments blinking”
(target weight selection mode) or
“The last two digits blinking”
(value adjustment mode).
RE-ZERO
key (To increase the value)
MODE
key
(To decrease the value)
To select the target weight or
adjust the value.
key
To store the new weight value.
Even if the AC adapter is
removed, the data is maintained
in non-volatile memory.
CAL
key
To cancel the operation and
return to
CC 0
.