33
5 Confirm that there is nothing on the pan and press the
key. The balance measures the zero point
and displays the measured value. Do not allow
vibration or drafts to affect the balance.
The balance displays the target weight value.
6 Place the displayed target weight on the pan and press
the
key. The balance measures the target
weight and displays the measured value. Do not allow
vibration or drafts to affect the balance.
7 The balance displays
end
. Remove the weight from
the pan.
8 When the GLP output is set, the balance displays
glp
1
and outputs “Calibration Test Report”.
9 The balance will automatically return to the weighing
mode.
From previous page
Target weight
GLP output