9-5 Description of the Item “Data Output Mode”
The parameter setting of the “Data output mode (
prt
)” applies to the performance when the data is
transmitted using the RS-232C serial interface.
Key mode
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the
weight data and the display blinks one time.
Required setting
dout
prt 0
Key
mode
Auto print modes A and B
When the displayed value is stable and the conditions of “Auto print polarity”, “Auto print difference” and
reference value are met, the balance outputs the weight data.
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the data
and the display blinks one time.
Mode A: Required setting
dout prt
1
Auto print mode A (reference = zero)
dout ap-p
Auto print polarity
dout ap-b
Auto print difference
Example
“For outputting the weight value each time a sample is added, with
“
ar-d
” set to “
1
” (to adjust zero after the data is output).”
Mode B
:
Required setting
dout prt
2
Auto print mode B (reference = last stable value)
dout ap-p
Auto print polarity
dout ap-b
Auto print difference
Example
“For outputting the weight value while a sample is added.”
Stream mode
The balance outputs the weight data continuously regardless of the display condition. The display does
not blink in this mode.
Required setting
dout prt
3
Stream
mode
ba5fnc 5pd
Display refresh rate
5if bp5
Baud
rate
Example
“For monitoring data on a computer”
Caution
The balance may not transmit the data completely at the specified refresh rate and baud rate. Set
the baud rate higher.
Key mode B
When the PRINT key is pressed, the balance outputs the weight data, regardless of the display
condition. The display does not blink in this mode.
Key mode C
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the
weight data. When the stabilization indicator is not on, the balance waits for the indicator to turn on and
outputs the data. The balance blinks one time in this mode.