56
14. CONNECTION TO PERIPHERAL EQUIPMENT
14-1 Connection to the AD-8121B Printer
Set the following parameters to use the AD-8121B printer.
Example of use
AD-8121B mode setting
To print A&D standard format weight data, using the FX-
i
WP
PRINT key
or FX-
i
WP
auto print mode. (The time and date can be added.)
MODE 1
To print A&D standard format weight data, using the AD-8121B
DATA key
or AD-8121B built-in timer. (The time and date can be added.)
To print, using the AD-8121B chart printing function.
MODE 2
To print the FX-
i
WP
statistical data.
MODE 3
To print GLP output.
MODE 3
Class
Item and
Parameter
Factory
settings
AD-8121B
MODE 1
AD-8121B
MODE 2
AD-8121B
MODE 3
prt
Data output mode
0
0
,
1
,
2
,
4
,
5
*1
3
0
,
1
,
2
,
4
,
5
*1
dout
Data output
pU5e
Data output pause
0
0
0
0
,
1
*2
bp5
Baud rate
2
2
2
2
btpr
Data bit, parity bit
0
0
0
0
Crlf
Terminator
0
0
0
0
5if
Serial interface
type
Data format
0
0
0
1
*1
Set appropriate parameters for “
ap-p
(Auto print polarity)” and “
ap-b
(Auto print difference) “ when
auto print mode A or B (
prt 1
or
2
) is selected.
Set the AD-8121B DIP switch No.3 to ON when unstable data is printed with “
prt 4
“.
*2
Set
1
when multiple lines are printed in the FX-
i
WP statistical calculation mode.
Notes
Refer to "10-2 GLP Report" for print samples.
Settings of AD-8121B DIP switches
MODE AD-8121B DIP switch
Description
MODE 1
1 2 3
4
MODE
ON
Print at receiving data.
Standard mode, statistical calculation mode
MODE 2
1 2 3
4
MODE
ON
Print by the AD-8121B DATA key operation or AD-8121B built-in timer.
Standard mode, interval mode, chart mode
MODE 3
1 2 3
4
MODE
ON
Print at receiving data.
Dump print mode
DIP switch No.3 : Handling unstable data
ON
OFF
Not
printed
Set the DIP switch No.4 to OFF.
1 2
3 4
MODE
ON
1 2
3 4
MODE
ON