19
With the zero point setting, tare and weighing range
About the operation at when power is turned on
The balance will determine the reference zero point when the ON:OFF key is pressed to enter the
weighing mode.
Depending on the load condition at that time, it will automatically judge whether to perform zero or tare
operation. The condition for determining which is used is "power on zero range", and when power on
zero range is exceeded, the tare subtraction operation is performed.
About re-zero operation
By pressing the RE-ZERO key, the display can be changed to zero.
Re-zero with the RE-ZERO key will automatically determine whether zero or tare operation is
performed.
The condition for determining which is used is "zero range", and when zero range is exceeded, the tare
subtraction operation is performed.
About measurement range
For the balance, the range that can be weighed is determined by model.
The total amount (net tare quantity) up to the maximum display of each model is displayed,
and when the maximum display is exceeded, E is displayed to indicate that the weighing range is
exceeded.
When in excess in negative, -E is displayed.
Model
Power on zero range
Zero range
-E display range
FZ-120
i
, FX-120
i
FZ-120
i
WP, FX-120
i
WP
FZ-120GD, FX-120GD
Approx.
±
60g
Approx.
±
2g
Approx.-60g or less
FZ-200
i,
FX-200
i
FZ-200
i
WP, FX-200
i
WP
FZ-200GD
,
FX-200GD
Approx.
±
4g
FZ-300
i,
FX-300
i
FZ-300
i
WP, FX-300
i
WP
FZ-300GD
,
FX-300GD
Approx.
±
6g
FZ-500
i,
FX-500
i
Approx.
±
10g
FZ-1200
i,
FX-1200
i
FZ-1200
i
WP, FX-1200
i
WP
FZ-1200GD, FX-1200GD
Approx.
±
600g
Approx.
±
20g
Approx.-600g or less
FZ-2000
i,
FX-2000
i
FZ-2000
i
WP, FX-2000
i
WP
FZ-2000GD
,
FX-2000GD
Approx.
±
40g
FZ-3000
i,
FX-3000
i
FZ-3000
i
WP, FX-3000
i
WP
FZ-3000GD
,
FX-3000GD
Approx.
±
60g
FZ-5000
i,
FX-5000
i
Approx.
±
100g
Summary of Contents for FZ-i WP Series
Page 95: ......