44
Class
Item and Parameter
Description
5if
Serial
interface
bp5
Baud rate
0
600 bps
1
1200 bps
2
2400 bps
3
4800 bps
4
9600 bps
5
19200 bps
btpr
Data bit, parity bit
0
7 bits, even
1
7 bits, odd
2
8 bits, none
Crlf
Terminator
0
CR LF
CR: ASCII code 0Dh
LF: ASCII code 0Ah
1
CR
type
Data format
0
A&D standard format
Refer to “10.6. Description of
the Item “Data Format””.
1
DP format
2
KF format
3
MT format
4
NU format
5
CSV format
t-Up
Timeout
0
No limit
Selects the wait time to receive
a command.
1
1 second
erCd
AK, Error code
0
No output
AK:ASCII code 06h
1
Output
mlt
Programmable-unit (Multi-unit)
Sets an arbitrary coefficient.
Available only when
programmable- unit mode is
selected.
Unit
Unit
Refer to “5. WEIGHING UNITS”.
Iid
ID number
Refer to “11. ID NUMBER AND GLP REPORT”
A
ap fnc
Application
apf
Application function
0
Normal weighing mode
1
Capacity indicator
2
Statistical calculation mode
5taf
Statistical function
mode output items
0
Number of data, sum
1
Number of data, sum, maximum, minimum, average, range
(maximum-minimum)
2
Number of data, sum, maximum, minimum, average, range
(maximum-minimum), standard deviation, coefficient of variation
3
Number of data, sum, maximum, minimum, average, range
(maximum-minimum), standard deviation, coefficient of
variation, relative error
locfnc
pa55
0
OFF
Refer to “14. PASSWORD
LOCK FUNCTION”.
1
ON (limits weighing operations)
2
ON (enables basic weighing)
loc no.
adm
IN
Administrator password input
U5er
01
~
U5er
10
User 1 password input
User 10 password input
*
: Only for the FZ-
i
/
FZ-
i
WP
/
FZ-GD series : Factory settings.
Summary of Contents for FZ-i WP Series
Page 95: ......